Blood Pressure Monitor/Electrocardiograph

rev.1.2

Manual
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0001Systolic blood pressure value0x5003kPa-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5003kPa-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0xF002Days of menstrual period0x0000NoneThe measurement value is fixed to 0. (The value does not have any meaning.)
HBP-9030Series
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x0010Mean Arterial Pressure0x5010mmHgonly for BLP (STP)
0x0011Pulse Rate Range Detection Flag0x0000Noneonly for BLP (STP)
0: Pulse rate is within the range
1: Pulse rate exceeds upper limit
2: Pulse rate is less than lower limit
0x0012Measurement Positon Detection Flag0x0000Noneonly for BLP (STP)
0: Proper measurement position
1: Improper measurement position
HEM-1026T2-AJC
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0004Posture (Result of Heart Zone Indicator)0x0000None0: Heart Zone Indicator OFF
1: Heart Zone OK
2: Heart Zone NG (too high)
3: Heart Zone NG (too low)
-2147483648(INT_MIN): Invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-1026T2-AJE
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0004Posture (Result of Heart Zone Indicator)0x0000None0: Heart Zone Indicator OFF
1: Heart Zone OK
2: Heart Zone NG (too high)
3: Heart Zone NG (too low)
-2147483648(INT_MIN): Invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-1026T2-AKA
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0004Posture (Result of Heart Zone Indicator)0x0000None0: Heart Zone Indicator OFF
1: Heart Zone OK
2: Heart Zone NG (too high)
3: Heart Zone NG (too low)
-2147483648(INT_MIN): Invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-6161T-E
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-6161T-RU
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-6161T2-BR
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-6231T-SH
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0004Posture (Result of Heart Zone Indicator)0x0000None0: Heart Zone Indicator OFF
1: Heart Zone OK
2: Heart Zone NG (too high)
3: Heart Zone NG (too low)
-2147483648(INT_MIN): Invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-6231T2-JC
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0004Posture (Result of Heart Zone Indicator)0x0000None0: Heart Zone Indicator OFF
1: Heart Zone OK
2: Heart Zone NG (too high)
3: Heart Zone NG (too low)
-2147483648(INT_MIN): Invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-6231T2-JE
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0004Posture (Result of Heart Zone Indicator)0x0000None0: Heart Zone Indicator OFF
1: Heart Zone OK
2: Heart Zone NG (too high)
3: Heart Zone NG (too low)
-2147483648(INT_MIN): Invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-6231T2-JT3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0004Posture (Result of Heart Zone Indicator)0x0000None0: Heart Zone Indicator OFF
1: Heart Zone OK
2: Heart Zone NG (too high)
3: Heart Zone NG (too low)
-2147483648(INT_MIN): Invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-6232T-AP
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0004Posture (Result of Heart Zone Indicator)0x0000None0: Heart Zone Indicator OFF
1: Heart Zone OK
2: Heart Zone NG (too high)
3: Heart Zone NG (too low)
-2147483648(INT_MIN): Invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-6232T-E
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0004Posture (Result of Heart Zone Indicator)0x0000None0: Heart Zone Indicator OFF
1: Heart Zone OK
2: Heart Zone NG (too high)
3: Heart Zone NG (too low)
-2147483648(INT_MIN): Invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-6233T
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0004Posture (Result of Heart Zone Indicator)0x0000None0: Heart Zone Indicator OFF
1: Heart Zone OK
2: Heart Zone NG (too high)
3: Heart Zone NG (too low)
-2147483648(INT_MIN): Invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-6320T-SH
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0004Posture (Result of Heart Zone Indicator)0x0000None0: Heart Zone Indicator OFF
1: Heart Zone OK
2: Heart Zone NG (too high)
3: Heart Zone NG (too low)
-2147483648(INT_MIN): Invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
HEM-6322T-SH
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0004Posture (Result of Heart Zone Indicator)0x0000None0: Heart Zone Indicator OFF
1: Heart Zone OK
2: Heart Zone NG (too high)
3: Heart Zone NG (too low)
-2147483648(INT_MIN): Invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
HEM-6323T
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0004Posture (Result of Heart Zone Indicator)0x0000None0: Heart Zone Indicator OFF
1: Heart Zone OK
2: Heart Zone NG (too high)
3: Heart Zone NG (too low)
-2147483648(INT_MIN): Invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
HEM-6324T
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0004Posture (Result of Heart Zone Indicator)0x0000None0: Heart Zone Indicator OFF
1: Heart Zone OK
2: Heart Zone NG (too high)
3: Heart Zone NG (too low)
-2147483648(INT_MIN): Invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
HEM-6325T
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0004Posture (Result of Heart Zone Indicator)0x0000None0: Heart Zone Indicator OFF
1: Heart Zone OK
2: Heart Zone NG (too high)
3: Heart Zone NG (too low)
-2147483648(INT_MIN): Invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
HEM-7136T-SH3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7138JT-SH
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7138K-SH
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7138T-SH
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7139T-SH3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7140T1-AP
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7141T1-AP
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7142T1-AP
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7142T2-AP
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7142T2_JAZ
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7143T1-AIN
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7143T1-AP
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7143T1_D
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7143T1_EBK
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7143T2_ESL
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7144T1-AU
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7144T2-BR
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7144T2-LA
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7146T2-JD
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7146T2-JF
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7150T-Z
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0025Irregular heartbeat detection count (displayed on LCD of main unit)0x0000NoneIrregular heartbeat detection count to be displayed on the LCD of main unit.
0~1 : Detection count is not displayed (controlled by the application).
2~3 : Detection count is displayed.
HEM-7151T-Z
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0025Irregular heartbeat detection count (displayed on LCD of main unit)0x0000NoneIrregular heartbeat detection count to be displayed on the LCD of main unit.
0~1 : Detection count is not displayed (controlled by the application).
2~3 : Detection count is displayed.
HEM-7153JT_ASH
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7153T_ASH
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7155T-ALRU
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7155T-D
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7155T-EBK
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7155T_AP
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7155T_ASH3BK
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7155T_ASH3SL
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7155T_ESL
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7155T_K4-ESL
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7156T-BR
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0025Irregular heartbeat detection count (displayed on LCD of main unit)0x0000NoneIrregular heartbeat detection count to be displayed on the LCD of main unit.
0~1 : Detection count is not displayed (controlled by the application).
2~3 : Detection count is displayed.
HEM-7156T-LA
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0025Irregular heartbeat detection count (displayed on LCD of main unit)0x0000NoneIrregular heartbeat detection count to be displayed on the LCD of main unit.
0~1 : Detection count is not displayed (controlled by the application).
2~3 : Detection count is displayed.
HEM-7156T_AAP
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7156T_AP
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7157T-AP
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7158T-JC
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7158T_AP3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7159T_AP3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-716DT2-LA
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0025Irregular heartbeat detection count (displayed on LCD of main unit)0x0000NoneIrregular heartbeat detection count to be displayed on the LCD of main unit.
0~1 : Detection count is not displayed (controlled by the application).
2~3 : Detection count is displayed.
HEM-7183T1-AP
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7183T1-CAP
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7183T1_LAP
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7194T1-FLAP
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0023Possible Afib detection flag0x0000None0 : not detected
1 : detected
HEM-7194T1-FLCAP
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0023Possible Afib detection flag0x0000None0 : not detected
1 : detected
HEM-7271L-SH3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7271P-SH3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7271T
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7271T_SH3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7280T-AP
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7280T-E
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7280T_TI-D
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7280T_TI-E
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7281T
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7282T
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7321T-ZV
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7322T-D
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7322T-E
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7325T
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7340T-Z
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0025Irregular heartbeat detection count (displayed on LCD of main unit)0x0000NoneIrregular heartbeat detection count to be displayed on the LCD of main unit.
0~1 : Detection count is not displayed (controlled by the application).
2~3 : Detection count is displayed.
HEM-7341T-Z
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0025Irregular heartbeat detection count (displayed on LCD of main unit)0x0000NoneIrregular heartbeat detection count to be displayed on the LCD of main unit.
0~1 : Detection count is not displayed (controlled by the application).
2~3 : Detection count is displayed.
HEM-7342T-Z
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0025Irregular heartbeat detection count (displayed on LCD of main unit)0x0000NoneIrregular heartbeat detection count to be displayed on the LCD of main unit.
0~1 : Detection count is not displayed (controlled by the application).
2~3 : Detection count is displayed.
HEM-7343T-Z
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0025Irregular heartbeat detection count (displayed on LCD of main unit)0x0000NoneIrregular heartbeat detection count to be displayed on the LCD of main unit.
0~1 : Detection count is not displayed (controlled by the application).
2~3 : Detection count is displayed.
HEM-7344JT_ASH3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7344T_ASH3BK
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7344T_ASH3SL
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7346T-AJC3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7346T-AJE3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7346T2-AJC32
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7346T2-AJE32
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7346T_ABR3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0025Irregular heartbeat detection count (displayed on LCD of main unit)0x0000NoneIrregular heartbeat detection count to be displayed on the LCD of main unit.
0~1 : Detection count is not displayed (controlled by the application).
2~3 : Detection count is displayed.
HEM-7346T_AP3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7347T-AJC3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7347T-AJE3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7347T2-AJC32
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7347T2-AJE32
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7349T_ABR
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0025Irregular heartbeat detection count (displayed on LCD of main unit)0x0000NoneIrregular heartbeat detection count to be displayed on the LCD of main unit.
0~1 : Detection count is not displayed (controlled by the application).
2~3 : Detection count is displayed.
HEM-7361T-ALRU
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0023Possible Afib detection flag0x0000None0: not detected
1: detected
0x0026Measurement mode0x0000NoneSee Measurement Mode
HEM-7361T-AP
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0023Possible Afib detection flag0x0000None0: not detected
1: detected
0x0026Measurement mode0x0000NoneSee Measurement Mode
HEM-7361T-D
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0023Possible Afib detection flag0x0000None0: not detected
1: detected
0x0026Measurement mode0x0000NoneSee Measurement Mode
HEM-7361T-EBK
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0023Possible Afib detection flag0x0000None0: not detected
1: detected
0x0026Measurement mode0x0000NoneSee Measurement Mode
HEM-7361T_ESL
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0023Possible Afib detection flag0x0000None0: not detected
1: detected
0x0026Measurement mode0x0000NoneSee Measurement Mode
HEM-7380T1-EBK
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0023Possible Afib detection flag0x0000None0: not detected
1: detected
HEM-7383T1-AP
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0023Possible Afib detection flag0x0000None0: not detected
1: detected
HEM-7384T1-NBBR
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0023Possible Afib detection flag0x0000None0: not detected
1: detected
HEM-7385T1-AJAZ3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0033Irregular pulse Interval detection Flag0x0000None0 : not detected
1 : detected
HEM-7386T1-AJF3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0033Irregular pulse Interval detection Flag0x0000None0 : not detected
1 : detected
HEM-7387T1-AJAZ3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0033Irregular pulse Interval detection Flag0x0000None0 : not detected
1 : detected
HEM-7388T1-AJF3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0033Irregular pulse Interval detection Flag0x0000None0 : not detected
1 : detected
HEM-7511T
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0005Room temperature0x3000
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-7530T1-BR3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x1002Average heart rate0xF0A0bpm
0x1012File type0x0000None0 : no digital filtering process (Original)
1 : reserve
2 : digital filtering process (Enhanced)
3: for development validation
Device setting information on measurement for SakuraOne/HCG-8010T.
0x1013Analysis result type and Analysis result0x0000NoneThe type number of analysis result is saved on Value while the text of analysis result is saved on DATA_TEXT.
200/201/207/214/218/296/297/298/299 for SakuraOne/KardiaMobile/KardiaMobile 6L.

For details, please refer to the 'ECG Analysis Result Type Details' sheet.
0x1014ECG monitoring data classification of UUID0x0000NoneUUID (string) generated by AliveCorSDK for every monitoring data
0x1015Kardia AI lib version0x0000NoneLibrary version for ECG analysis of AliveCorSDK
(obtained from ACKEcgEvaluation)
0x1016ECG sampling rate0xF0E0HzFrequency of sampling when recording ECG waveform
Fixed to 125 for HCG-8010T
0x1017Commercial power frequency0xF0E0Hz50 : 50Hz
60 : 60Hz
Device setting information during recording
0x1018Setting for number of leads0x0000None1 : single lead
6 : six leads
Fixed to 1 for SakuraOne/KardiaMobile/HCG-8010T
0x1019Maximum recording time0xF050Second30/60/120/180/240/300
Device setting information at the time of recording (maximum number of seconds for recording)
Fixed to 30 for HCG-8010T
0x101AActual recording time0xF050SecondNumber of seconds for actual recording of ECG waveform
For HCG-8010T, measurement time is calculated from the number of data entries
0x101BFile path of waveform data (originalPath)0x0000NoneFile path for waveform data (atc file) in the filter type of Original.
Shared by iOS and Android, the file path is set directly under the directory for saving atc files.
0x101CFile path of waveform data (enhancedPath)0x0000NoneFile path for waveform data (atc file) in the filter type of Enhanced.
Shared by iOS and Android, the file path is set directly under the directory for saving atc files.
0x101DECG recording device type0x0000NoneDevice type of ECG recording
0 : Not set
1 : SakuraOne
2 : KardiaMobile
3 : KardiaMobile 6L
4: HCG-8010T
0x102FAverage heart rate (for displaying)0xF0A0bpmNo limitation for SakuraOne/KardiaMobile.
* If the value is outside the supported range:
-2147483648 (INT_MIN)
0x1081Measurement part0x0000None0 : Not entered (unknown)
1 : chest
2 : wrist
3 : wrist and ankle (induction: 6-Lead of KM6L)
11: chest (V1)
12: chest (V2)
13: chest (V3)
14: chest (V4)
15: chest (V5)
16: chest (V6)
17: abdomen
18: palm
* No unit as this is a user-entered item.

Fixed to 2 for SakuraOne/KardiaMobile.
For KardiaMobile 6L, 2 for 1 lead and 3 for 6 leads. 11–18 are for HCG-8010T only
0x1082Symptom0x0000NoneInclusive OR of all items corresponding to selected symptoms is saved to Value. The input text for symptoms is saved in DATA_TEXT if 'others' is selected.
Bitwise inclusive OR of the following symptoms.
0x01: none
0x02: others
0x04: chest pain
0x08: dizziness
0x10: fatigue
0x20: nausea
0x40: palpitation
0x80: shortness of breath
* If the value is 0, it will be considered as 'no input'.
* No unit as this is a user-entered item.
0x1083Immediately preceding intensity of exercise0x0000None0 : not entered
1 : high
2 : medium
3 : low
4 : resting
* No unit as this is a user-entered item.
0x1084Note (for ECG recording)0x0000NoneNote by user when ECG waveform is recorded.
* No unit as this is a user-entered item.
0x1085ECG waveform graph flipped upside down0x0000None0 : no flipping
1 : flip upside down
* No flipping other than 1
The value that is set in the details of waveform graph. The default value is 0.
0x1087ECG waveform graph with filter settings0x0000None0 : no digital filtering process (Original)
1 : reserve
2 : with digital filtering process (Enhanced)
The value that is set in the details of waveform graph.
For SakuraOne/KardiaMobile, the default value is the one set during measurement on a filter-type device.
HEM-7530T_AP3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x1002Average heart rate0xF0A0bpm
0x1012File type0x0000None0 : no digital filtering process (Original)
1 : reserve
2 : digital filtering process (Enhanced)
3: for development validation
Device setting information on measurement for SakuraOne/HCG-8010T.
0x1013Analysis result type and Analysis result0x0000NoneThe type number of analysis result is saved on Value while the text of analysis result is saved on DATA_TEXT.
200/201/207/214/218/296/297/298/299 for SakuraOne/KardiaMobile/KardiaMobile 6L.

For details, please refer to the 'ECG Analysis Result Type Details' sheet.
0x1014ECG monitoring data classification of UUID0x0000NoneUUID (string) generated by AliveCorSDK for every monitoring data
0x1015Kardia AI lib version0x0000NoneLibrary version for ECG analysis of AliveCorSDK
(obtained from ACKEcgEvaluation)
0x1016ECG sampling rate0xF0E0HzFrequency of sampling when recording ECG waveform
Fixed to 125 for HCG-8010T
0x1017Commercial power frequency0xF0E0Hz50 : 50Hz
60 : 60Hz
Device setting information during recording
0x1018Setting for number of leads0x0000None1 : single lead
6 : six leads
Fixed to 1 for SakuraOne/KardiaMobile/HCG-8010T
0x1019Maximum recording time0xF050Second30/60/120/180/240/300
Device setting information at the time of recording (maximum number of seconds for recording)
Fixed to 30 for HCG-8010T
0x101AActual recording time0xF050SecondNumber of seconds for actual recording of ECG waveform
For HCG-8010T, measurement time is calculated from the number of data entries
0x101BFile path of waveform data (originalPath)0x0000NoneFile path for waveform data (atc file) in the filter type of Original.
Shared by iOS and Android, the file path is set directly under the directory for saving atc files.
0x101CFile path of waveform data (enhancedPath)0x0000NoneFile path for waveform data (atc file) in the filter type of Enhanced.
Shared by iOS and Android, the file path is set directly under the directory for saving atc files.
0x101DECG recording device type0x0000NoneDevice type of ECG recording
0 : Not set
1 : SakuraOne
2 : KardiaMobile
3 : KardiaMobile 6L
4: HCG-8010T
0x102FAverage heart rate (for displaying)0xF0A0bpmNo limitation for SakuraOne/KardiaMobile.
* If the value is outside the supported range:
-2147483648 (INT_MIN)
0x1081Measurement part0x0000None0 : Not entered (unknown)
1 : chest
2 : wrist
3 : wrist and ankle (induction: 6-Lead of KM6L)
11: chest (V1)
12: chest (V2)
13: chest (V3)
14: chest (V4)
15: chest (V5)
16: chest (V6)
17: abdomen
18: palm
* No unit as this is a user-entered item.

Fixed to 2 for SakuraOne/KardiaMobile.
For KardiaMobile 6L, 2 for 1 lead and 3 for 6 leads. 11–18 are for HCG-8010T only
0x1082Symptom0x0000NoneInclusive OR of all items corresponding to selected symptoms is saved to Value. The input text for symptoms is saved in DATA_TEXT if 'others' is selected.
Bitwise inclusive OR of the following symptoms.
0x01: none
0x02: others
0x04: chest pain
0x08: dizziness
0x10: fatigue
0x20: nausea
0x40: palpitation
0x80: shortness of breath
* If the value is 0, it will be considered as 'no input'.
* No unit as this is a user-entered item.
0x1083Immediately preceding intensity of exercise0x0000None0 : not entered
1 : high
2 : medium
3 : low
4 : resting
* No unit as this is a user-entered item.
0x1084Note (for ECG recording)0x0000NoneNote by user when ECG waveform is recorded.
* No unit as this is a user-entered item.
0x1085ECG waveform graph flipped upside down0x0000None0 : no flipping
1 : flip upside down
* No flipping other than 1
The value that is set in the details of waveform graph. The default value is 0.
0x1087ECG waveform graph with filter settings0x0000None0 : no digital filtering process (Original)
1 : reserve
2 : with digital filtering process (Enhanced)
The value that is set in the details of waveform graph.
For SakuraOne/KardiaMobile, the default value is the one set during measurement on a filter-type device.
HEM-7530T_E3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x1002Average heart rate0xF0A0bpm
0x1012File type0x0000None0 : no digital filtering process (Original)
1 : reserve
2 : digital filtering process (Enhanced)
3: for development validation
Device setting information on measurement for SakuraOne/HCG-8010T.
0x1013Analysis result type and Analysis result0x0000NoneThe type number of analysis result is saved on Value while the text of analysis result is saved on DATA_TEXT.
200/201/207/214/218/296/297/298/299 for SakuraOne/KardiaMobile/KardiaMobile 6L.

For details, please refer to the 'ECG Analysis Result Type Details' sheet.
0x1014ECG monitoring data classification of UUID0x0000NoneUUID (string) generated by AliveCorSDK for every monitoring data
0x1015Kardia AI lib version0x0000NoneLibrary version for ECG analysis of AliveCorSDK
(obtained from ACKEcgEvaluation)
0x1016ECG sampling rate0xF0E0HzFrequency of sampling when recording ECG waveform
Fixed to 125 for HCG-8010T
0x1017Commercial power frequency0xF0E0Hz50 : 50Hz
60 : 60Hz
Device setting information during recording
0x1018Setting for number of leads0x0000None1 : single lead
6 : six leads
Fixed to 1 for SakuraOne/KardiaMobile/HCG-8010T
0x1019Maximum recording time0xF050Second30/60/120/180/240/300
Device setting information at the time of recording (maximum number of seconds for recording)
Fixed to 30 for HCG-8010T
0x101AActual recording time0xF050SecondNumber of seconds for actual recording of ECG waveform
For HCG-8010T, measurement time is calculated from the number of data entries
0x101BFile path of waveform data (originalPath)0x0000NoneFile path for waveform data (atc file) in the filter type of Original.
Shared by iOS and Android, the file path is set directly under the directory for saving atc files.
0x101CFile path of waveform data (enhancedPath)0x0000NoneFile path for waveform data (atc file) in the filter type of Enhanced.
Shared by iOS and Android, the file path is set directly under the directory for saving atc files.
0x101DECG recording device type0x0000NoneDevice type of ECG recording
0 : Not set
1 : SakuraOne
2 : KardiaMobile
3 : KardiaMobile 6L
4: HCG-8010T
0x102FAverage heart rate (for displaying)0xF0A0bpmNo limitation for SakuraOne/KardiaMobile.
* If the value is outside the supported range:
-2147483648 (INT_MIN)
0x1081Measurement part0x0000None0 : Not entered (unknown)
1 : chest
2 : wrist
3 : wrist and ankle (induction: 6-Lead of KM6L)
11: chest (V1)
12: chest (V2)
13: chest (V3)
14: chest (V4)
15: chest (V5)
16: chest (V6)
17: abdomen
18: palm
* No unit as this is a user-entered item.

Fixed to 2 for SakuraOne/KardiaMobile.
For KardiaMobile 6L, 2 for 1 lead and 3 for 6 leads. 11–18 are for HCG-8010T only
0x1082Symptom0x0000NoneInclusive OR of all items corresponding to selected symptoms is saved to Value. The input text for symptoms is saved in DATA_TEXT if 'others' is selected.
Bitwise inclusive OR of the following symptoms.
0x01: none
0x02: others
0x04: chest pain
0x08: dizziness
0x10: fatigue
0x20: nausea
0x40: palpitation
0x80: shortness of breath
* If the value is 0, it will be considered as 'no input'.
* No unit as this is a user-entered item.
0x1083Immediately preceding intensity of exercise0x0000None0 : not entered
1 : high
2 : medium
3 : low
4 : resting
* No unit as this is a user-entered item.
0x1084Note (for ECG recording)0x0000NoneNote by user when ECG waveform is recorded.
* No unit as this is a user-entered item.
0x1085ECG waveform graph flipped upside down0x0000None0 : no flipping
1 : flip upside down
* No flipping other than 1
The value that is set in the details of waveform graph. The default value is 0.
0x1087ECG waveform graph with filter settings0x0000None0 : no digital filtering process (Original)
1 : reserve
2 : with digital filtering process (Enhanced)
The value that is set in the details of waveform graph.
For SakuraOne/KardiaMobile, the default value is the one set during measurement on a filter-type device.
HEM-7530T_J3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x1002Average heart rate0xF0A0bpm
0x1012File type0x0000None0 : no digital filtering process (Original)
1 : reserve
2 : digital filtering process (Enhanced)
3: for development validation
Device setting information on measurement for SakuraOne/HCG-8010T.
0x1013Analysis result type and Analysis result0x0000NoneThe type number of analysis result is saved on Value while the text of analysis result is saved on DATA_TEXT.
200/201/207/214/218/296/297/298/299 for SakuraOne/KardiaMobile/KardiaMobile 6L.

For details, please refer to the 'ECG Analysis Result Type Details' sheet.
0x1014ECG monitoring data classification of UUID0x0000NoneUUID (string) generated by AliveCorSDK for every monitoring data
0x1015Kardia AI lib version0x0000NoneLibrary version for ECG analysis of AliveCorSDK
(obtained from ACKEcgEvaluation)
0x1016ECG sampling rate0xF0E0HzFrequency of sampling when recording ECG waveform
Fixed to 125 for HCG-8010T
0x1017Commercial power frequency0xF0E0Hz50 : 50Hz
60 : 60Hz
Device setting information during recording
0x1018Setting for number of leads0x0000None1 : single lead
6 : six leads
Fixed to 1 for SakuraOne/KardiaMobile/HCG-8010T
0x1019Maximum recording time0xF050Second30/60/120/180/240/300
Device setting information at the time of recording (maximum number of seconds for recording)
Fixed to 30 for HCG-8010T
0x101AActual recording time0xF050SecondNumber of seconds for actual recording of ECG waveform
For HCG-8010T, measurement time is calculated from the number of data entries
0x101BFile path of waveform data (originalPath)0x0000NoneFile path for waveform data (atc file) in the filter type of Original.
Shared by iOS and Android, the file path is set directly under the directory for saving atc files.
0x101CFile path of waveform data (enhancedPath)0x0000NoneFile path for waveform data (atc file) in the filter type of Enhanced.
Shared by iOS and Android, the file path is set directly under the directory for saving atc files.
0x101DECG recording device type0x0000NoneDevice type of ECG recording
0 : Not set
1 : SakuraOne
2 : KardiaMobile
3 : KardiaMobile 6L
4: HCG-8010T
0x102FAverage heart rate (for displaying)0xF0A0bpmNo limitation for SakuraOne/KardiaMobile.
* If the value is outside the supported range:
-2147483648 (INT_MIN)
0x1081Measurement part0x0000None0 : Not entered (unknown)
1 : chest
2 : wrist
3 : wrist and ankle (induction: 6-Lead of KM6L)
11: chest (V1)
12: chest (V2)
13: chest (V3)
14: chest (V4)
15: chest (V5)
16: chest (V6)
17: abdomen
18: palm
* No unit as this is a user-entered item.

Fixed to 2 for SakuraOne/KardiaMobile.
For KardiaMobile 6L, 2 for 1 lead and 3 for 6 leads. 11–18 are for HCG-8010T only
0x1082Symptom0x0000NoneInclusive OR of all items corresponding to selected symptoms is saved to Value. The input text for symptoms is saved in DATA_TEXT if 'others' is selected.
Bitwise inclusive OR of the following symptoms.
0x01: none
0x02: others
0x04: chest pain
0x08: dizziness
0x10: fatigue
0x20: nausea
0x40: palpitation
0x80: shortness of breath
* If the value is 0, it will be considered as 'no input'.
* No unit as this is a user-entered item.
0x1083Immediately preceding intensity of exercise0x0000None0 : not entered
1 : high
2 : medium
3 : low
4 : resting
* No unit as this is a user-entered item.
0x1084Note (for ECG recording)0x0000NoneNote by user when ECG waveform is recorded.
* No unit as this is a user-entered item.
0x1085ECG waveform graph flipped upside down0x0000None0 : no flipping
1 : flip upside down
* No flipping other than 1
The value that is set in the details of waveform graph. The default value is 0.
0x1087ECG waveform graph with filter settings0x0000None0 : no digital filtering process (Original)
1 : reserve
2 : with digital filtering process (Enhanced)
The value that is set in the details of waveform graph.
For SakuraOne/KardiaMobile, the default value is the one set during measurement on a filter-type device.
HEM-7530T_JT3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x1002Average heart rate0xF0A0bpm
0x1012File type0x0000None0 : no digital filtering process (Original)
1 : reserve
2 : digital filtering process (Enhanced)
3: for development validation
Device setting information on measurement for SakuraOne/HCG-8010T.
0x1013Analysis result type and Analysis result0x0000NoneThe type number of analysis result is saved on Value while the text of analysis result is saved on DATA_TEXT.
200/201/207/214/218/296/297/298/299 for SakuraOne/KardiaMobile/KardiaMobile 6L.

For details, please refer to the 'ECG Analysis Result Type Details' sheet.
0x1014ECG monitoring data classification of UUID0x0000NoneUUID (string) generated by AliveCorSDK for every monitoring data
0x1015Kardia AI lib version0x0000NoneLibrary version for ECG analysis of AliveCorSDK
(obtained from ACKEcgEvaluation)
0x1016ECG sampling rate0xF0E0HzFrequency of sampling when recording ECG waveform
Fixed to 125 for HCG-8010T
0x1017Commercial power frequency0xF0E0Hz50 : 50Hz
60 : 60Hz
Device setting information during recording
0x1018Setting for number of leads0x0000None1 : single lead
6 : six leads
Fixed to 1 for SakuraOne/KardiaMobile/HCG-8010T
0x1019Maximum recording time0xF050Second30/60/120/180/240/300
Device setting information at the time of recording (maximum number of seconds for recording)
Fixed to 30 for HCG-8010T
0x101AActual recording time0xF050SecondNumber of seconds for actual recording of ECG waveform
For HCG-8010T, measurement time is calculated from the number of data entries
0x101BFile path of waveform data (originalPath)0x0000NoneFile path for waveform data (atc file) in the filter type of Original.
Shared by iOS and Android, the file path is set directly under the directory for saving atc files.
0x101CFile path of waveform data (enhancedPath)0x0000NoneFile path for waveform data (atc file) in the filter type of Enhanced.
Shared by iOS and Android, the file path is set directly under the directory for saving atc files.
0x101DECG recording device type0x0000NoneDevice type of ECG recording
0 : Not set
1 : SakuraOne
2 : KardiaMobile
3 : KardiaMobile 6L
4: HCG-8010T
0x102FAverage heart rate (for displaying)0xF0A0bpmNo limitation for SakuraOne/KardiaMobile.
* If the value is outside the supported range:
-2147483648 (INT_MIN)
0x1081Measurement part0x0000None0 : Not entered (unknown)
1 : chest
2 : wrist
3 : wrist and ankle (induction: 6-Lead of KM6L)
11: chest (V1)
12: chest (V2)
13: chest (V3)
14: chest (V4)
15: chest (V5)
16: chest (V6)
17: abdomen
18: palm
* No unit as this is a user-entered item.

Fixed to 2 for SakuraOne/KardiaMobile.
For KardiaMobile 6L, 2 for 1 lead and 3 for 6 leads. 11–18 are for HCG-8010T only
0x1082Symptom0x0000NoneInclusive OR of all items corresponding to selected symptoms is saved to Value. The input text for symptoms is saved in DATA_TEXT if 'others' is selected.
Bitwise inclusive OR of the following symptoms.
0x01: none
0x02: others
0x04: chest pain
0x08: dizziness
0x10: fatigue
0x20: nausea
0x40: palpitation
0x80: shortness of breath
* If the value is 0, it will be considered as 'no input'.
* No unit as this is a user-entered item.
0x1083Immediately preceding intensity of exercise0x0000None0 : not entered
1 : high
2 : medium
3 : low
4 : resting
* No unit as this is a user-entered item.
0x1084Note (for ECG recording)0x0000NoneNote by user when ECG waveform is recorded.
* No unit as this is a user-entered item.
0x1085ECG waveform graph flipped upside down0x0000None0 : no flipping
1 : flip upside down
* No flipping other than 1
The value that is set in the details of waveform graph. The default value is 0.
0x1087ECG waveform graph with filter settings0x0000None0 : no digital filtering process (Original)
1 : reserve
2 : with digital filtering process (Enhanced)
The value that is set in the details of waveform graph.
For SakuraOne/KardiaMobile, the default value is the one set during measurement on a filter-type device.
HEM-7600T
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
HEM-7600T-E
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
HEM-7600T-SH3BK
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
HEM-7600T2-JF
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
HEM-7600T_W
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
HEM-7600T_W-SH3W
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
HEM-7600T_W-Z
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
HEM-8630T-SH
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0004Posture
(Result of Heart Zone Indicator)
0x0000None0 : Heart Zone Indicator OFF
1 : Heart Zone OK
2 : Heart Zone NG (too high)
3 : Heart Zone NG (too low)
-2147483648(INT_MIN): Invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-8732K-SH
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-8732T-SH
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
HEM-9601T-J3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0004Posture (Result of Heart Zone Indicator)0x0000None0: Heart Zone Indicator OFF
1: Heart Zone OK
2: Heart Zone NG (too high)
3: Heart Zone NG (too low)
-2147483648(INT_MIN): Invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0026Measurement mode0x0000Nonerefer to 'Measurement Mode'
HEM-9601T2-BR3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0004Posture (Result of Heart Zone Indicator)0x0000None0: Heart Zone Indicator OFF
1: Heart Zone OK
2: Heart Zone NG (too high)
3: Heart Zone NG (too low)
-2147483648(INT_MIN): Invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0026Measurement mode0x0000Nonerefer to 'Measurement Mode'
HEM-9601T_E3
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0004Posture (Result of Heart Zone Indicator)0x0000None0: Heart Zone Indicator OFF
1: Heart Zone OK
2: Heart Zone NG (too high)
3: Heart Zone NG (too low)
-2147483648(INT_MIN): Invalid
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0026Measurement mode0x0000Nonerefer to 'Measurement Mode'
HEM-9700T
ID HexNameUnit ID HexUnit NameDescription
0x0001Systolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0002Diastolic blood pressure value0x5010mmHg-2147483648 (INT_MIN): invalid
0x0003Pulse rate0xF0A0bpm-2147483648 (INT_MIN): invalid
0x0005Room temperature0x3000
0x0006Irregular pulse rate detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0007Body movement detection flag0x0000None0: not detected
1: detected
-2147483648(INT_MIN): Invalid
0x0008Cuff Wrap Guide check0x0000None0: loosely wrapped
1: securely wrapped
0x000AConsecutive measurement count0x0000None0: Single measurement
1: 1st TruRead measurement
2: 2nd TruRead measurement
3: 3rd TruRead measurement
Fixed to 0 if TruRead function is disabled.
0x0026Measurement mode0x0000Nonerefer to 'Measurement Mode'
HCG-8010T1-J
ID HexNameUnit ID HexUnit NameDescription
0x1000Maximum heart rate0xF0A0bpm
0x1001Minimum heart rate0xF0A0bpm
0x1002Average heart rate0xF0A0bpm
0x1003Average value of instant heart rate0xF0A0bpm
0x1010Number of 125 Hz ECG waveform data0x0000None
0x1011125 Hz ECG waveform data0x0000None+1023 ~ -1024 [*9.757uV]
0x1012File type0x0000None0 : no digital filtering process (Original)
1 : reserve
2 : digital filtering process (Enhanced)
3: for development validation
Device setting information on measurement for SakuraOne/HCG-8010T.
0x1013Analysis result type and Analysis result0x0000None"The type number of analysis result is saved on Value while the text of analysis result is saved on DATA_TEXT.
200/201/207/214/218/296/297/298/299 for SakuraOne/KardiaMobile/KardiaMobile 6L.

For details, please refer to the 'ECG Analysis Result Type Details' sheet.
0x1014ECG monitoring data classification of UUID0x0000NoneUUID (string) generated by AliveCorSDK for every monitoring data
0x1016ECG sampling rate0xF0E0HzFrequency of sampling when recording ECG waveform
Fixed to 125 for HCG-8010T
0x1018Setting for number of leads0x0000None1 : single lead
6 : six leads
Fixed to 1 for SakuraOne/KardiaMobile/HCG-8010T
0x1019Maximum recording time0xF050Second30/60/120/180/240/300
Device setting information at the time of recording (maximum number of seconds for recording)
Fixed to 30 for HCG-8010T
0x101AActual recording time0xF050SecondNumber of seconds for actual recording of ECG waveform
For HCG-8010T, measurement time is calculated from the number of data entries
0x101DECG recording device type0x0000NoneDevice type of ECG recording
0 : Not set
1 : SakuraOne
2 : KardiaMobile
3 : KardiaMobile 6L
4: HCG-8010T
0x101FArrhythmia event information (for storage)0x0000None
0x1020Arrhythmia event information flag (Pause)0x0000None0:false, 1:true
0x1021Arrhythmia event information flag (V-RUN)0x0000None0:false, 1:true
0x1022Arrhythmia event information flag (Couplet)0x0000None0:false, 1:true
0x1023Arrhythmia event information flag (VPC)0x0000None0:false, 1:true
0x1024Arrhythmia event information flag (S-RUN)0x0000None0:false, 1:true
0x1025Arrhythmia event information flag (Tachy)0x0000None0:false, 1:true
0x1026Arrhythmia event information flag (Brady)0x0000None0:false, 1:true
0x1027Arrhythmia event information flag (SVPC)0x0000None0:false, 1:true
0x1028Arrhythmia event information flag (ST-T abnormality)0x0000None0:false, 1:true
0x1029Arrhythmia event information flag (irregular RR interval)0x0000None0:false, 1:true
0x102BArrhythmia event information flag (NOISE)0x0000None0:false, 1:true
0x102CArrhythmia event information flag (others)0x0000None0:false, 1:true
0x102DArrhythmia event information flag (no average waveform)0x0000None0:false, 1:true
0x102EArrhythmia event information flag (no RR fluctuation rate)0x0000None0:false, 1:true
0x102FAverage heart rate (for displaying)0xF0A0bpmNo limitation for SakuraOne/KardiaMobile.
* If the value is outside the supported range:
-2147483648 (INT_MIN)
0x1030Analysis result flag0x0000NoneHCG-8010T only
Refer to the sheet 'Supplement: HCG-8010T Analysis Result Display Specification'
0x1031Data format version0x0000NoneIdentification information to check the data storage format
*Unsigned 4-byte
0x1032Event analysis module version0x0000NoneNumber to check the version of event analysis module
*Unsigned short 2-byte
0x1033Time post-correction flag0x0000NoneFlag used for time post-correction function.
0: Time not set
1: Time set
2: Time corrected
3: Not subject to time correction
0x1034Contact determination information0x0000NoneFlag indicating the result of contact determination.
*Set the result of the following bit operations.
b0: Contact OK
b1: Weak contact of finger electrode
b2: Unstable contact of finger electrode
b3: Weak contact of chest electrode
b4: Unstable contact of chest electrode
b5: Reserved
0x1081Measurement part0x0000None0 : Not entered (unknown)
1 : chest
2 : wrist
3 : wrist and ankle (induction: 6-Lead of KM6L)
11: chest (V1)
12: chest (V2)
13: chest (V3)
14: chest (V4)
15: chest (V5)
16: chest (V6)
17: abdomen
18: palm
* No unit as this is a user-entered item.

Fixed to 2 for SakuraOne/KardiaMobile.
For KardiaMobile 6L, 2 for 1 lead and 3 for 6 leads. 11–18 are for HCG-8010T only
0x1084Note (for ECG recording)0x0000NoneNote by user when ECG waveform is recorded.
* No unit as this is a user-entered item.
0x1086ECG waveform graph with sensitivity settings0x0000None1 : 5mm/mV
2 : 0mm/mV (default)
3 : 20mm/mV
* 0 : not defined (process as 10mm/mV)
0x1088Measurement status
HCG-8010T only
0x0000NoneSelected measurement status is saved in Value, and input symptom text is saved in DATA_TEXT when 'others' is selected.
0x0000: not selected
0x0002: others
0x0004: upon waking
0x0008: after eating/drinking
0x0010: at rest
0x0020: after using toilet
0x0040: after smoking
0x0080: after bathing
0x0100: after exercise
0x0200: after drinking alcohol
0x0400: after taking medication
* No unit as it is a user input item
* '0x0001: none' is treated as missing number
0x1089Measurement posture
HCG-8010T only
0x0000NoneSelected measurement posture is saved in Value, and input symptom text is saved in DATA_TEXT when 'others' is selected.
0x0000: not selected
0x0002: others
0x0004: sitting posture
0x0008: standing posture
0x0010: supine posture
* No unit as it is a user input item
* '0x0001: none' is treated as missing number
0x108ASubjective symptoms
For HCG-8010T
0x0000NoneLogical OR of selected symptom items is saved in Value, and input symptom text is saved in DATA_TEXT when 'others' is selected.
Logical OR of the following symptom bits:
0x0001: none
0x0002: others
0x0004: chest pain
0x0008: dizziness
0x0010: fatigue
0x0020: nausea
0x0040: palpitation
0x0080: shortness of breath
0x0100: chest pressure
0x0200: pricking chest pain
0x0400: difficulty breathing
0x0800: lightheadedness
0x1000: fainting
0x2000: skipped heartbeat
* If the value is 0, it is treated as not entered
* No unit as it is a user input item
* Separate symptom input indicator IDs for SakuraOne/KardiaMobile6L and HCG-8010T (may be unified in future; define item for 0x1082 as well)
HCG-8011B-J
ID HexNameUnit ID HexUnit NameDescription
0x1002Average heart rate0xF0A0bpm
0x1012File type0x0000None0 : no digital filtering process (Original)
1 : reserve
2 : digital filtering process (Enhanced)
3: for development validation
Device setting information on measurement for SakuraOne/HCG-8010T.
0x1013Analysis result type and Analysis result0x0000None"The type number of analysis result is saved on Value while the text of analysis result is saved on DATA_TEXT.
200/201/207/214/218/296/297/298/299 for SakuraOne/KardiaMobile/KardiaMobile 6L.

For details, please refer to the 'ECG Analysis Result Type Details' sheet.
0x1014ECG monitoring data classification of UUID0x0000NoneUUID (string) generated by AliveCorSDK for every monitoring data
0x1015Kardia AI lib version0x0000NoneLibrary version for ECG analysis of AliveCorSDK
(obtained from ACKEcgEvaluation)
0x1016ECG sampling rate0xF0E0HzFrequency of sampling when recording ECG waveform Fixed to 125 for HCG-8010T
0x1017Commercial power frequency0xF0E0Hz50 : 50Hz
60 : 60Hz
Device setting information during recording
0x1018Setting for number of leads0x0000None1 : single lead
6 : six leads
Fixed to 1 for SakuraOne/KardiaMobile/HCG-8010T
0x1019Maximum recording time0xF050Second30/60/120/180/240/300
Device setting information at the time of recording (maximum number of seconds for recording)
Fixed to 30 for HCG-8010T
0x101AActual recording time0xF050SecondNumber of seconds for actual recording of ECG waveform
For HCG-8010T, measurement time is calculated from the number of data entries
0x101BFile path of waveform data (originalPath)0x0000NoneFile path for waveform data (atc file) in the filter type of Original.
Shared by iOS and Android, the file path is set directly under the directory for saving atc files.
0x101CFile path of waveform data (enhancedPath)0x0000NoneFile path for waveform data (atc file) in the filter type of Enhanced.
Shared by iOS and Android, the file path is set directly under the directory for saving atc files.
0x101DECG recording device type0x0000NoneDevice type of ECG recording
0 : Not set
1 : SakuraOne
2 : KardiaMobile
3 : KardiaMobile 6L
4: HCG-8010T
0x102FAverage heart rate (for displaying)0xF0A0bpmNo limitation for SakuraOne/KardiaMobile.
* If the value is outside the supported range:
-2147483648 (INT_MIN)
0x1081Measurement part0x0000None0 : Not entered (unknown)
1 : chest
2 : wrist
3 : wrist and ankle (induction: 6-Lead of KM6L)
11: chest (V1)
12: chest (V2)
13: chest (V3)
14: chest (V4)
15: chest (V5)
16: chest (V6)
17: abdomen
18: palm
* No unit as this is a user-entered item.

Fixed to 2 for SakuraOne/KardiaMobile.
For KardiaMobile 6L, 2 for 1 lead and 3 for 6 leads. 11–18 are for HCG-8010T only
0x1082Symptom0x0000NoneInclusive OR of all items corresponding to selected symptoms is saved to Value. The input text for symptoms is saved in DATA_TEXT if 'others' is selected.
Bitwise inclusive OR of the following symptoms.
0x01: none
0x02: others
0x04: chest pain
0x08: dizziness
0x10: fatigue
0x20: nausea
0x40: palpitation
0x80: shortness of breath
* If the value is 0, it will be considered as 'no input'.
* No unit as this is a user-entered item.
0x1083Immediately preceding intensity of exercise0x0000None0 : not entered
1 : high
2 : medium
3 : low
4 : resting
* No unit as this is a user-entered item.
0x1084Note (for ECG recording)0x0000NoneNote by user when ECG waveform is recorded.
* No unit as this is a user-entered item.
0x1085ECG waveform graph flipped upside down0x0000None0 : no flipping
1 : flip upside down
* No flipping other than 1
The value that is set in the details of waveform graph. The default value is 0.
0x1087ECG waveform graph with filter settings0x0000None0 : no digital filtering process (Original)
1 : reserve
2 : with digital filtering process (Enhanced)
The value that is set in the details of waveform graph.
For SakuraOne/KardiaMobile, the default value is the one set during measurement on a filter-type device.
HCG-8060T
ID HexNameUnit ID HexUnit NameDescription
0x1002Average heart rate0xF0A0bpm
0x1012File type0x0000None0 : no digital filtering process (Original)
1 : reserve
2 : digital filtering process (Enhanced)
3: for development validation
Device setting information on measurement for SakuraOne/HCG-8010T.
0x1013Analysis result type and Analysis result0x0000None"The type number of analysis result is saved on Value while the text of analysis result is saved on DATA_TEXT.
200/201/207/214/218/296/297/298/299 for SakuraOne/KardiaMobile/KardiaMobile 6L.

For details, please refer to the 'ECG Analysis Result Type Details' sheet.
0x1014ECG monitoring data classification of UUID0x0000NoneUUID (string) generated by AliveCorSDK for every monitoring data
0x1015Kardia AI lib version0x0000NoneLibrary version for ECG analysis of AliveCorSDK
(obtained from ACKEcgEvaluation)
0x1016ECG sampling rate0xF0E0HzFrequency of sampling when recording ECG waveform Fixed to 125 for HCG-8010T
0x1017Commercial power frequency0xF0E0Hz50 : 50Hz
60 : 60Hz
Device setting information during recording
0x1018Setting for number of leads0x0000None1 : single lead
6 : six leads
Fixed to 1 for SakuraOne/KardiaMobile/HCG-8010T
0x1019Maximum recording time0xF050Second30/60/120/180/240/300
Device setting information at the time of recording (maximum number of seconds for recording)
Fixed to 30 for HCG-8010T
0x101AActual recording time0xF050SecondNumber of seconds for actual recording of ECG waveform
For HCG-8010T, measurement time is calculated from the number of data entries
0x101BFile path of waveform data (originalPath)0x0000NoneFile path for waveform data (atc file) in the filter type of Original.
Shared by iOS and Android, the file path is set directly under the directory for saving atc files.
0x101CFile path of waveform data (enhancedPath)0x0000NoneFile path for waveform data (atc file) in the filter type of Enhanced.
Shared by iOS and Android, the file path is set directly under the directory for saving atc files.
0x101DECG recording device type0x0000NoneDevice type of ECG recording
0 : Not set
1 : SakuraOne
2 : KardiaMobile
3 : KardiaMobile 6L
4: HCG-8010T
0x102FAverage heart rate (for displaying)0xF0A0bpmNo limitation for SakuraOne/KardiaMobile.
* If the value is outside the supported range:
-2147483648 (INT_MIN)
0x1081Measurement part0x0000None0 : Not entered (unknown)
1 : chest
2 : wrist
3 : wrist and ankle (induction: 6-Lead of KM6L)
11: chest (V1)
12: chest (V2)
13: chest (V3)
14: chest (V4)
15: chest (V5)
16: chest (V6)
17: abdomen
18: palm
* No unit as this is a user-entered item.

Fixed to 2 for SakuraOne/KardiaMobile.
For KardiaMobile 6L, 2 for 1 lead and 3 for 6 leads. 11–18 are for HCG-8010T only
0x1082Symptom0x0000NoneInclusive OR of all items corresponding to selected symptoms is saved to Value. The input text for symptoms is saved in DATA_TEXT if 'others' is selected.
Bitwise inclusive OR of the following symptoms.
0x01: none
0x02: others
0x04: chest pain
0x08: dizziness
0x10: fatigue
0x20: nausea
0x40: palpitation
0x80: shortness of breath
* If the value is 0, it will be considered as 'no input'.
* No unit as this is a user-entered item.
0x1083Immediately preceding intensity of exercise0x0000None0 : not entered
1 : high
2 : medium
3 : low
4 : resting
* No unit as this is a user-entered item.
0x1084Note (for ECG recording)0x0000NoneNote by user when ECG waveform is recorded.
* No unit as this is a user-entered item.
0x1085ECG waveform graph flipped upside down0x0000None0 : no flipping
1 : flip upside down
* No flipping other than 1
The value that is set in the details of waveform graph. The default value is 0.
0x1087ECG waveform graph with filter settings0x0000None0 : no digital filtering process (Original)
1 : reserve
2 : with digital filtering process (Enhanced)
The value that is set in the details of waveform graph.
For SakuraOne/KardiaMobile, the default value is the one set during measurement on a filter-type device.
HCG-8060T_GL
ID HexNameUnit ID HexUnit NameDescription
0x1002Average heart rate0xF0A0bpm
0x1012File type0x0000None0 : no digital filtering process (Original)
1 : reserve
2 : digital filtering process (Enhanced)
3: for development validation
Device setting information on measurement for SakuraOne/HCG-8010T.
0x1013Analysis result type and Analysis result0x0000None"The type number of analysis result is saved on Value while the text of analysis result is saved on DATA_TEXT.
200/201/207/214/218/296/297/298/299 for SakuraOne/KardiaMobile/KardiaMobile 6L.

For details, please refer to the 'ECG Analysis Result Type Details' sheet.
0x1014ECG monitoring data classification of UUID0x0000NoneUUID (string) generated by AliveCorSDK for every monitoring data
0x1015Kardia AI lib version0x0000NoneLibrary version for ECG analysis of AliveCorSDK
(obtained from ACKEcgEvaluation)
0x1016ECG sampling rate0xF0E0HzFrequency of sampling when recording ECG waveform Fixed to 125 for HCG-8010T
0x1017Commercial power frequency0xF0E0Hz50 : 50Hz
60 : 60Hz
Device setting information during recording
0x1018Setting for number of leads0x0000None1 : single lead
6 : six leads
Fixed to 1 for SakuraOne/KardiaMobile/HCG-8010T
0x1019Maximum recording time0xF050Second30/60/120/180/240/300
Device setting information at the time of recording (maximum number of seconds for recording)
Fixed to 30 for HCG-8010T
0x101AActual recording time0xF050SecondNumber of seconds for actual recording of ECG waveform
For HCG-8010T, measurement time is calculated from the number of data entries
0x101BFile path of waveform data (originalPath)0x0000NoneFile path for waveform data (atc file) in the filter type of Original.
Shared by iOS and Android, the file path is set directly under the directory for saving atc files.
0x101CFile path of waveform data (enhancedPath)0x0000NoneFile path for waveform data (atc file) in the filter type of Enhanced.
Shared by iOS and Android, the file path is set directly under the directory for saving atc files.
0x101DECG recording device type0x0000NoneDevice type of ECG recording
0 : Not set
1 : SakuraOne
2 : KardiaMobile
3 : KardiaMobile 6L
4: HCG-8010T
0x102FAverage heart rate (for displaying)0xF0A0bpmNo limitation for SakuraOne/KardiaMobile.
* If the value is outside the supported range:
-2147483648 (INT_MIN)
0x1081Measurement part0x0000None0 : Not entered (unknown)
1 : chest
2 : wrist
3 : wrist and ankle (induction: 6-Lead of KM6L)
11: chest (V1)
12: chest (V2)
13: chest (V3)
14: chest (V4)
15: chest (V5)
16: chest (V6)
17: abdomen
18: palm
* No unit as this is a user-entered item.

Fixed to 2 for SakuraOne/KardiaMobile.
For KardiaMobile 6L, 2 for 1 lead and 3 for 6 leads. 11–18 are for HCG-8010T only
0x1082Symptom0x0000NoneInclusive OR of all items corresponding to selected symptoms is saved to Value. The input text for symptoms is saved in DATA_TEXT if 'others' is selected.
Bitwise inclusive OR of the following symptoms.
0x01: none
0x02: others
0x04: chest pain
0x08: dizziness
0x10: fatigue
0x20: nausea
0x40: palpitation
0x80: shortness of breath
* If the value is 0, it will be considered as 'no input'.
* No unit as this is a user-entered item.
0x1083Immediately preceding intensity of exercise0x0000None0 : not entered
1 : high
2 : medium
3 : low
4 : resting
* No unit as this is a user-entered item.
0x1084Note (for ECG recording)0x0000NoneNote by user when ECG waveform is recorded.
* No unit as this is a user-entered item.
0x1085ECG waveform graph flipped upside down0x0000None0 : no flipping
1 : flip upside down
* No flipping other than 1
The value that is set in the details of waveform graph. The default value is 0.
0x1087ECG waveform graph with filter settings0x0000None0 : no digital filtering process (Original)
1 : reserve
2 : with digital filtering process (Enhanced)
The value that is set in the details of waveform graph.
For SakuraOne/KardiaMobile, the default value is the one set during measurement on a filter-type device.
KM-6L
ID HexNameUnit ID HexUnit NameDescription
0x1002Average heart rate0xF0A0bpm
0x1012File type0x0000None0 : no digital filtering process (Original)
1 : reserve
2 : digital filtering process (Enhanced)
3: for development validation
Device setting information on measurement for SakuraOne/HCG-8010T.
0x1013Analysis result type and Analysis result0x0000None"The type number of analysis result is saved on Value while the text of analysis result is saved on DATA_TEXT.
200/201/207/214/218/296/297/298/299 for SakuraOne/KardiaMobile/KardiaMobile 6L.

For details, please refer to the 'ECG Analysis Result Type Details' sheet.
0x1014ECG monitoring data classification of UUID0x0000NoneUUID (string) generated by AliveCorSDK for every monitoring data
0x1015Kardia AI lib version0x0000NoneLibrary version for ECG analysis of AliveCorSDK
(obtained from ACKEcgEvaluation)
0x1016ECG sampling rate0xF0E0HzFrequency of sampling when recording ECG waveform Fixed to 125 for HCG-8010T
0x1017Commercial power frequency0xF0E0Hz50 : 50Hz
60 : 60Hz
Device setting information during recording
0x1018Setting for number of leads0x0000None1 : single lead
6 : six leads
Fixed to 1 for SakuraOne/KardiaMobile/HCG-8010T
0x1019Maximum recording time0xF050Second30/60/120/180/240/300
Device setting information at the time of recording (maximum number of seconds for recording)
Fixed to 30 for HCG-8010T
0x101AActual recording time0xF050SecondNumber of seconds for actual recording of ECG waveform
For HCG-8010T, measurement time is calculated from the number of data entries
0x101BFile path of waveform data (originalPath)0x0000NoneFile path for waveform data (atc file) in the filter type of Original.
Shared by iOS and Android, the file path is set directly under the directory for saving atc files.
0x101CFile path of waveform data (enhancedPath)0x0000NoneFile path for waveform data (atc file) in the filter type of Enhanced.
Shared by iOS and Android, the file path is set directly under the directory for saving atc files.
0x101DECG recording device type0x0000NoneDevice type of ECG recording
0 : Not set
1 : SakuraOne
2 : KardiaMobile
3 : KardiaMobile 6L
4: HCG-8010T
0x102FAverage heart rate (for displaying)0xF0A0bpmNo limitation for SakuraOne/KardiaMobile.
* If the value is outside the supported range:
-2147483648 (INT_MIN)
0x1081Measurement part0x0000None0 : Not entered (unknown)
1 : chest
2 : wrist
3 : wrist and ankle (induction: 6-Lead of KM6L)
11: chest (V1)
12: chest (V2)
13: chest (V3)
14: chest (V4)
15: chest (V5)
16: chest (V6)
17: abdomen
18: palm
* No unit as this is a user-entered item.

Fixed to 2 for SakuraOne/KardiaMobile.
For KardiaMobile 6L, 2 for 1 lead and 3 for 6 leads. 11–18 are for HCG-8010T only
0x1082Symptom0x0000NoneInclusive OR of all items corresponding to selected symptoms is saved to Value. The input text for symptoms is saved in DATA_TEXT if 'others' is selected.
Bitwise inclusive OR of the following symptoms.
0x01: none
0x02: others
0x04: chest pain
0x08: dizziness
0x10: fatigue
0x20: nausea
0x40: palpitation
0x80: shortness of breath
* If the value is 0, it will be considered as 'no input'.
* No unit as this is a user-entered item.
0x1083Immediately preceding intensity of exercise0x0000None0 : not entered
1 : high
2 : medium
3 : low
4 : resting
* No unit as this is a user-entered item.
0x1084Note (for ECG recording)0x0000NoneNote by user when ECG waveform is recorded.
* No unit as this is a user-entered item.
0x1085ECG waveform graph flipped upside down0x0000None0 : no flipping
1 : flip upside down
* No flipping other than 1
The value that is set in the details of waveform graph. The default value is 0.
0x1087ECG waveform graph with filter settings0x0000None0 : no digital filtering process (Original)
1 : reserve
2 : with digital filtering process (Enhanced)
The value that is set in the details of waveform graph.
For SakuraOne/KardiaMobile, the default value is the one set during measurement on a filter-type device.