Blood Pressure Monitor/Electrocardiograph
rev.1.2
Manual
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0001 | Systolic blood pressure value | 0x5003 | kPa | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5003 | kPa | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0xF002 | Days of menstrual period | 0x0000 | None | The measurement value is fixed to 0. (The value does not have any meaning.) |
HBP-9030Series
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x0010 | Mean Arterial Pressure | 0x5010 | mmHg | only for BLP (STP) |
| 0x0011 | Pulse Rate Range Detection Flag | 0x0000 | None | only for BLP (STP) 0: Pulse rate is within the range 1: Pulse rate exceeds upper limit 2: Pulse rate is less than lower limit |
| 0x0012 | Measurement Positon Detection Flag | 0x0000 | None | only for BLP (STP) 0: Proper measurement position 1: Improper measurement position |
HEM-1026T2-AJC
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0004 | Posture (Result of Heart Zone Indicator) | 0x0000 | None | 0: Heart Zone Indicator OFF 1: Heart Zone OK 2: Heart Zone NG (too high) 3: Heart Zone NG (too low) -2147483648(INT_MIN): Invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-1026T2-AJE
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0004 | Posture (Result of Heart Zone Indicator) | 0x0000 | None | 0: Heart Zone Indicator OFF 1: Heart Zone OK 2: Heart Zone NG (too high) 3: Heart Zone NG (too low) -2147483648(INT_MIN): Invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-1026T2-AKA
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0004 | Posture (Result of Heart Zone Indicator) | 0x0000 | None | 0: Heart Zone Indicator OFF 1: Heart Zone OK 2: Heart Zone NG (too high) 3: Heart Zone NG (too low) -2147483648(INT_MIN): Invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-6161T-E
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-6161T-RU
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-6161T2-BR
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-6231T-SH
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0004 | Posture (Result of Heart Zone Indicator) | 0x0000 | None | 0: Heart Zone Indicator OFF 1: Heart Zone OK 2: Heart Zone NG (too high) 3: Heart Zone NG (too low) -2147483648(INT_MIN): Invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-6231T2-JC
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0004 | Posture (Result of Heart Zone Indicator) | 0x0000 | None | 0: Heart Zone Indicator OFF 1: Heart Zone OK 2: Heart Zone NG (too high) 3: Heart Zone NG (too low) -2147483648(INT_MIN): Invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-6231T2-JE
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0004 | Posture (Result of Heart Zone Indicator) | 0x0000 | None | 0: Heart Zone Indicator OFF 1: Heart Zone OK 2: Heart Zone NG (too high) 3: Heart Zone NG (too low) -2147483648(INT_MIN): Invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-6231T2-JT3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0004 | Posture (Result of Heart Zone Indicator) | 0x0000 | None | 0: Heart Zone Indicator OFF 1: Heart Zone OK 2: Heart Zone NG (too high) 3: Heart Zone NG (too low) -2147483648(INT_MIN): Invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-6232T-AP
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0004 | Posture (Result of Heart Zone Indicator) | 0x0000 | None | 0: Heart Zone Indicator OFF 1: Heart Zone OK 2: Heart Zone NG (too high) 3: Heart Zone NG (too low) -2147483648(INT_MIN): Invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-6232T-E
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0004 | Posture (Result of Heart Zone Indicator) | 0x0000 | None | 0: Heart Zone Indicator OFF 1: Heart Zone OK 2: Heart Zone NG (too high) 3: Heart Zone NG (too low) -2147483648(INT_MIN): Invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-6233T
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0004 | Posture (Result of Heart Zone Indicator) | 0x0000 | None | 0: Heart Zone Indicator OFF 1: Heart Zone OK 2: Heart Zone NG (too high) 3: Heart Zone NG (too low) -2147483648(INT_MIN): Invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-6320T-SH
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0004 | Posture (Result of Heart Zone Indicator) | 0x0000 | None | 0: Heart Zone Indicator OFF 1: Heart Zone OK 2: Heart Zone NG (too high) 3: Heart Zone NG (too low) -2147483648(INT_MIN): Invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
HEM-6322T-SH
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0004 | Posture (Result of Heart Zone Indicator) | 0x0000 | None | 0: Heart Zone Indicator OFF 1: Heart Zone OK 2: Heart Zone NG (too high) 3: Heart Zone NG (too low) -2147483648(INT_MIN): Invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
HEM-6323T
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0004 | Posture (Result of Heart Zone Indicator) | 0x0000 | None | 0: Heart Zone Indicator OFF 1: Heart Zone OK 2: Heart Zone NG (too high) 3: Heart Zone NG (too low) -2147483648(INT_MIN): Invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
HEM-6324T
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0004 | Posture (Result of Heart Zone Indicator) | 0x0000 | None | 0: Heart Zone Indicator OFF 1: Heart Zone OK 2: Heart Zone NG (too high) 3: Heart Zone NG (too low) -2147483648(INT_MIN): Invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
HEM-6325T
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0004 | Posture (Result of Heart Zone Indicator) | 0x0000 | None | 0: Heart Zone Indicator OFF 1: Heart Zone OK 2: Heart Zone NG (too high) 3: Heart Zone NG (too low) -2147483648(INT_MIN): Invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
HEM-7136T-SH3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7138JT-SH
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7138K-SH
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7138T-SH
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7139T-SH3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7140T1-AP
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7141T1-AP
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7142T1-AP
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7142T2-AP
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7142T2_JAZ
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7143T1-AIN
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7143T1-AP
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7143T1_D
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7143T1_EBK
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7143T2_ESL
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7144T1-AU
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7144T2-BR
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7144T2-LA
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7146T2-JD
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7146T2-JF
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7150T-Z
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0025 | Irregular heartbeat detection count (displayed on LCD of main unit) | 0x0000 | None | Irregular heartbeat detection count to be displayed on the LCD of main unit. 0~1 : Detection count is not displayed (controlled by the application). 2~3 : Detection count is displayed. |
HEM-7151T-Z
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0025 | Irregular heartbeat detection count (displayed on LCD of main unit) | 0x0000 | None | Irregular heartbeat detection count to be displayed on the LCD of main unit. 0~1 : Detection count is not displayed (controlled by the application). 2~3 : Detection count is displayed. |
HEM-7153JT_ASH
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7153T_ASH
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7155T-ALRU
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7155T-D
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7155T-EBK
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7155T_AP
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7155T_ASH3BK
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7155T_ASH3SL
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7155T_ESL
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7155T_K4-ESL
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7156T-BR
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0025 | Irregular heartbeat detection count (displayed on LCD of main unit) | 0x0000 | None | Irregular heartbeat detection count to be displayed on the LCD of main unit. 0~1 : Detection count is not displayed (controlled by the application). 2~3 : Detection count is displayed. |
HEM-7156T-LA
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0025 | Irregular heartbeat detection count (displayed on LCD of main unit) | 0x0000 | None | Irregular heartbeat detection count to be displayed on the LCD of main unit. 0~1 : Detection count is not displayed (controlled by the application). 2~3 : Detection count is displayed. |
HEM-7156T_AAP
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7156T_AP
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7157T-AP
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7158T-JC
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7158T_AP3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7159T_AP3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-716DT2-LA
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0025 | Irregular heartbeat detection count (displayed on LCD of main unit) | 0x0000 | None | Irregular heartbeat detection count to be displayed on the LCD of main unit. 0~1 : Detection count is not displayed (controlled by the application). 2~3 : Detection count is displayed. |
HEM-7183T1-AP
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7183T1-CAP
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7183T1_LAP
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7194T1-FLAP
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0023 | Possible Afib detection flag | 0x0000 | None | 0 : not detected 1 : detected |
HEM-7194T1-FLCAP
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0023 | Possible Afib detection flag | 0x0000 | None | 0 : not detected 1 : detected |
HEM-7271L-SH3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7271P-SH3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7271T
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7271T_SH3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7280T-AP
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7280T-E
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7280T_TI-D
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7280T_TI-E
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7281T
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7282T
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7321T-ZV
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7322T-D
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7322T-E
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7325T
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7340T-Z
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0025 | Irregular heartbeat detection count (displayed on LCD of main unit) | 0x0000 | None | Irregular heartbeat detection count to be displayed on the LCD of main unit. 0~1 : Detection count is not displayed (controlled by the application). 2~3 : Detection count is displayed. |
HEM-7341T-Z
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0025 | Irregular heartbeat detection count (displayed on LCD of main unit) | 0x0000 | None | Irregular heartbeat detection count to be displayed on the LCD of main unit. 0~1 : Detection count is not displayed (controlled by the application). 2~3 : Detection count is displayed. |
HEM-7342T-Z
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0025 | Irregular heartbeat detection count (displayed on LCD of main unit) | 0x0000 | None | Irregular heartbeat detection count to be displayed on the LCD of main unit. 0~1 : Detection count is not displayed (controlled by the application). 2~3 : Detection count is displayed. |
HEM-7343T-Z
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0025 | Irregular heartbeat detection count (displayed on LCD of main unit) | 0x0000 | None | Irregular heartbeat detection count to be displayed on the LCD of main unit. 0~1 : Detection count is not displayed (controlled by the application). 2~3 : Detection count is displayed. |
HEM-7344JT_ASH3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7344T_ASH3BK
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7344T_ASH3SL
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7346T-AJC3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7346T-AJE3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7346T2-AJC32
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7346T2-AJE32
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7346T_ABR3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0025 | Irregular heartbeat detection count (displayed on LCD of main unit) | 0x0000 | None | Irregular heartbeat detection count to be displayed on the LCD of main unit. 0~1 : Detection count is not displayed (controlled by the application). 2~3 : Detection count is displayed. |
HEM-7346T_AP3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7347T-AJC3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7347T-AJE3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7347T2-AJC32
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7347T2-AJE32
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7349T_ABR
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0025 | Irregular heartbeat detection count (displayed on LCD of main unit) | 0x0000 | None | Irregular heartbeat detection count to be displayed on the LCD of main unit. 0~1 : Detection count is not displayed (controlled by the application). 2~3 : Detection count is displayed. |
HEM-7361T-ALRU
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0023 | Possible Afib detection flag | 0x0000 | None | 0: not detected 1: detected |
| 0x0026 | Measurement mode | 0x0000 | None | See Measurement Mode |
HEM-7361T-AP
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0023 | Possible Afib detection flag | 0x0000 | None | 0: not detected 1: detected |
| 0x0026 | Measurement mode | 0x0000 | None | See Measurement Mode |
HEM-7361T-D
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0023 | Possible Afib detection flag | 0x0000 | None | 0: not detected 1: detected |
| 0x0026 | Measurement mode | 0x0000 | None | See Measurement Mode |
HEM-7361T-EBK
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0023 | Possible Afib detection flag | 0x0000 | None | 0: not detected 1: detected |
| 0x0026 | Measurement mode | 0x0000 | None | See Measurement Mode |
HEM-7361T_ESL
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0023 | Possible Afib detection flag | 0x0000 | None | 0: not detected 1: detected |
| 0x0026 | Measurement mode | 0x0000 | None | See Measurement Mode |
HEM-7380T1-EBK
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0023 | Possible Afib detection flag | 0x0000 | None | 0: not detected 1: detected |
HEM-7383T1-AP
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0023 | Possible Afib detection flag | 0x0000 | None | 0: not detected 1: detected |
HEM-7384T1-NBBR
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0023 | Possible Afib detection flag | 0x0000 | None | 0: not detected 1: detected |
HEM-7385T1-AJAZ3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0033 | Irregular pulse Interval detection Flag | 0x0000 | None | 0 : not detected 1 : detected |
HEM-7386T1-AJF3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0033 | Irregular pulse Interval detection Flag | 0x0000 | None | 0 : not detected 1 : detected |
HEM-7387T1-AJAZ3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0033 | Irregular pulse Interval detection Flag | 0x0000 | None | 0 : not detected 1 : detected |
HEM-7388T1-AJF3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0033 | Irregular pulse Interval detection Flag | 0x0000 | None | 0 : not detected 1 : detected |
HEM-7511T
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0005 | Room temperature | 0x3000 | ℃ | |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-7530T1-BR3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x1002 | Average heart rate | 0xF0A0 | bpm | |
| 0x1012 | File type | 0x0000 | None | 0 : no digital filtering process (Original) 1 : reserve 2 : digital filtering process (Enhanced) 3: for development validation Device setting information on measurement for SakuraOne/HCG-8010T. |
| 0x1013 | Analysis result type and Analysis result | 0x0000 | None | The type number of analysis result is saved on Value while the text of analysis result is saved on DATA_TEXT. 200/201/207/214/218/296/297/298/299 for SakuraOne/KardiaMobile/KardiaMobile 6L. For details, please refer to the 'ECG Analysis Result Type Details' sheet. |
| 0x1014 | ECG monitoring data classification of UUID | 0x0000 | None | UUID (string) generated by AliveCorSDK for every monitoring data |
| 0x1015 | Kardia AI lib version | 0x0000 | None | Library version for ECG analysis of AliveCorSDK (obtained from ACKEcgEvaluation) |
| 0x1016 | ECG sampling rate | 0xF0E0 | Hz | Frequency of sampling when recording ECG waveform Fixed to 125 for HCG-8010T |
| 0x1017 | Commercial power frequency | 0xF0E0 | Hz | 50 : 50Hz 60 : 60Hz Device setting information during recording |
| 0x1018 | Setting for number of leads | 0x0000 | None | 1 : single lead 6 : six leads Fixed to 1 for SakuraOne/KardiaMobile/HCG-8010T |
| 0x1019 | Maximum recording time | 0xF050 | Second | 30/60/120/180/240/300 Device setting information at the time of recording (maximum number of seconds for recording) Fixed to 30 for HCG-8010T |
| 0x101A | Actual recording time | 0xF050 | Second | Number of seconds for actual recording of ECG waveform For HCG-8010T, measurement time is calculated from the number of data entries |
| 0x101B | File path of waveform data (originalPath) | 0x0000 | None | File path for waveform data (atc file) in the filter type of Original. Shared by iOS and Android, the file path is set directly under the directory for saving atc files. |
| 0x101C | File path of waveform data (enhancedPath) | 0x0000 | None | File path for waveform data (atc file) in the filter type of Enhanced. Shared by iOS and Android, the file path is set directly under the directory for saving atc files. |
| 0x101D | ECG recording device type | 0x0000 | None | Device type of ECG recording 0 : Not set 1 : SakuraOne 2 : KardiaMobile 3 : KardiaMobile 6L 4: HCG-8010T |
| 0x102F | Average heart rate (for displaying) | 0xF0A0 | bpm | No limitation for SakuraOne/KardiaMobile. * If the value is outside the supported range: -2147483648 (INT_MIN) |
| 0x1081 | Measurement part | 0x0000 | None | 0 : Not entered (unknown) 1 : chest 2 : wrist 3 : wrist and ankle (induction: 6-Lead of KM6L) 11: chest (V1) 12: chest (V2) 13: chest (V3) 14: chest (V4) 15: chest (V5) 16: chest (V6) 17: abdomen 18: palm * No unit as this is a user-entered item. Fixed to 2 for SakuraOne/KardiaMobile. For KardiaMobile 6L, 2 for 1 lead and 3 for 6 leads. 11–18 are for HCG-8010T only |
| 0x1082 | Symptom | 0x0000 | None | Inclusive OR of all items corresponding to selected symptoms is saved to Value. The input text for symptoms is saved in DATA_TEXT if 'others' is selected. Bitwise inclusive OR of the following symptoms. 0x01: none 0x02: others 0x04: chest pain 0x08: dizziness 0x10: fatigue 0x20: nausea 0x40: palpitation 0x80: shortness of breath * If the value is 0, it will be considered as 'no input'. * No unit as this is a user-entered item. |
| 0x1083 | Immediately preceding intensity of exercise | 0x0000 | None | 0 : not entered 1 : high 2 : medium 3 : low 4 : resting * No unit as this is a user-entered item. |
| 0x1084 | Note (for ECG recording) | 0x0000 | None | Note by user when ECG waveform is recorded. * No unit as this is a user-entered item. |
| 0x1085 | ECG waveform graph flipped upside down | 0x0000 | None | 0 : no flipping 1 : flip upside down * No flipping other than 1 The value that is set in the details of waveform graph. The default value is 0. |
| 0x1087 | ECG waveform graph with filter settings | 0x0000 | None | 0 : no digital filtering process (Original) 1 : reserve 2 : with digital filtering process (Enhanced) The value that is set in the details of waveform graph. For SakuraOne/KardiaMobile, the default value is the one set during measurement on a filter-type device. |
HEM-7530T_AP3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x1002 | Average heart rate | 0xF0A0 | bpm | |
| 0x1012 | File type | 0x0000 | None | 0 : no digital filtering process (Original) 1 : reserve 2 : digital filtering process (Enhanced) 3: for development validation Device setting information on measurement for SakuraOne/HCG-8010T. |
| 0x1013 | Analysis result type and Analysis result | 0x0000 | None | The type number of analysis result is saved on Value while the text of analysis result is saved on DATA_TEXT. 200/201/207/214/218/296/297/298/299 for SakuraOne/KardiaMobile/KardiaMobile 6L. For details, please refer to the 'ECG Analysis Result Type Details' sheet. |
| 0x1014 | ECG monitoring data classification of UUID | 0x0000 | None | UUID (string) generated by AliveCorSDK for every monitoring data |
| 0x1015 | Kardia AI lib version | 0x0000 | None | Library version for ECG analysis of AliveCorSDK (obtained from ACKEcgEvaluation) |
| 0x1016 | ECG sampling rate | 0xF0E0 | Hz | Frequency of sampling when recording ECG waveform Fixed to 125 for HCG-8010T |
| 0x1017 | Commercial power frequency | 0xF0E0 | Hz | 50 : 50Hz 60 : 60Hz Device setting information during recording |
| 0x1018 | Setting for number of leads | 0x0000 | None | 1 : single lead 6 : six leads Fixed to 1 for SakuraOne/KardiaMobile/HCG-8010T |
| 0x1019 | Maximum recording time | 0xF050 | Second | 30/60/120/180/240/300 Device setting information at the time of recording (maximum number of seconds for recording) Fixed to 30 for HCG-8010T |
| 0x101A | Actual recording time | 0xF050 | Second | Number of seconds for actual recording of ECG waveform For HCG-8010T, measurement time is calculated from the number of data entries |
| 0x101B | File path of waveform data (originalPath) | 0x0000 | None | File path for waveform data (atc file) in the filter type of Original. Shared by iOS and Android, the file path is set directly under the directory for saving atc files. |
| 0x101C | File path of waveform data (enhancedPath) | 0x0000 | None | File path for waveform data (atc file) in the filter type of Enhanced. Shared by iOS and Android, the file path is set directly under the directory for saving atc files. |
| 0x101D | ECG recording device type | 0x0000 | None | Device type of ECG recording 0 : Not set 1 : SakuraOne 2 : KardiaMobile 3 : KardiaMobile 6L 4: HCG-8010T |
| 0x102F | Average heart rate (for displaying) | 0xF0A0 | bpm | No limitation for SakuraOne/KardiaMobile. * If the value is outside the supported range: -2147483648 (INT_MIN) |
| 0x1081 | Measurement part | 0x0000 | None | 0 : Not entered (unknown) 1 : chest 2 : wrist 3 : wrist and ankle (induction: 6-Lead of KM6L) 11: chest (V1) 12: chest (V2) 13: chest (V3) 14: chest (V4) 15: chest (V5) 16: chest (V6) 17: abdomen 18: palm * No unit as this is a user-entered item. Fixed to 2 for SakuraOne/KardiaMobile. For KardiaMobile 6L, 2 for 1 lead and 3 for 6 leads. 11–18 are for HCG-8010T only |
| 0x1082 | Symptom | 0x0000 | None | Inclusive OR of all items corresponding to selected symptoms is saved to Value. The input text for symptoms is saved in DATA_TEXT if 'others' is selected. Bitwise inclusive OR of the following symptoms. 0x01: none 0x02: others 0x04: chest pain 0x08: dizziness 0x10: fatigue 0x20: nausea 0x40: palpitation 0x80: shortness of breath * If the value is 0, it will be considered as 'no input'. * No unit as this is a user-entered item. |
| 0x1083 | Immediately preceding intensity of exercise | 0x0000 | None | 0 : not entered 1 : high 2 : medium 3 : low 4 : resting * No unit as this is a user-entered item. |
| 0x1084 | Note (for ECG recording) | 0x0000 | None | Note by user when ECG waveform is recorded. * No unit as this is a user-entered item. |
| 0x1085 | ECG waveform graph flipped upside down | 0x0000 | None | 0 : no flipping 1 : flip upside down * No flipping other than 1 The value that is set in the details of waveform graph. The default value is 0. |
| 0x1087 | ECG waveform graph with filter settings | 0x0000 | None | 0 : no digital filtering process (Original) 1 : reserve 2 : with digital filtering process (Enhanced) The value that is set in the details of waveform graph. For SakuraOne/KardiaMobile, the default value is the one set during measurement on a filter-type device. |
HEM-7530T_E3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x1002 | Average heart rate | 0xF0A0 | bpm | |
| 0x1012 | File type | 0x0000 | None | 0 : no digital filtering process (Original) 1 : reserve 2 : digital filtering process (Enhanced) 3: for development validation Device setting information on measurement for SakuraOne/HCG-8010T. |
| 0x1013 | Analysis result type and Analysis result | 0x0000 | None | The type number of analysis result is saved on Value while the text of analysis result is saved on DATA_TEXT. 200/201/207/214/218/296/297/298/299 for SakuraOne/KardiaMobile/KardiaMobile 6L. For details, please refer to the 'ECG Analysis Result Type Details' sheet. |
| 0x1014 | ECG monitoring data classification of UUID | 0x0000 | None | UUID (string) generated by AliveCorSDK for every monitoring data |
| 0x1015 | Kardia AI lib version | 0x0000 | None | Library version for ECG analysis of AliveCorSDK (obtained from ACKEcgEvaluation) |
| 0x1016 | ECG sampling rate | 0xF0E0 | Hz | Frequency of sampling when recording ECG waveform Fixed to 125 for HCG-8010T |
| 0x1017 | Commercial power frequency | 0xF0E0 | Hz | 50 : 50Hz 60 : 60Hz Device setting information during recording |
| 0x1018 | Setting for number of leads | 0x0000 | None | 1 : single lead 6 : six leads Fixed to 1 for SakuraOne/KardiaMobile/HCG-8010T |
| 0x1019 | Maximum recording time | 0xF050 | Second | 30/60/120/180/240/300 Device setting information at the time of recording (maximum number of seconds for recording) Fixed to 30 for HCG-8010T |
| 0x101A | Actual recording time | 0xF050 | Second | Number of seconds for actual recording of ECG waveform For HCG-8010T, measurement time is calculated from the number of data entries |
| 0x101B | File path of waveform data (originalPath) | 0x0000 | None | File path for waveform data (atc file) in the filter type of Original. Shared by iOS and Android, the file path is set directly under the directory for saving atc files. |
| 0x101C | File path of waveform data (enhancedPath) | 0x0000 | None | File path for waveform data (atc file) in the filter type of Enhanced. Shared by iOS and Android, the file path is set directly under the directory for saving atc files. |
| 0x101D | ECG recording device type | 0x0000 | None | Device type of ECG recording 0 : Not set 1 : SakuraOne 2 : KardiaMobile 3 : KardiaMobile 6L 4: HCG-8010T |
| 0x102F | Average heart rate (for displaying) | 0xF0A0 | bpm | No limitation for SakuraOne/KardiaMobile. * If the value is outside the supported range: -2147483648 (INT_MIN) |
| 0x1081 | Measurement part | 0x0000 | None | 0 : Not entered (unknown) 1 : chest 2 : wrist 3 : wrist and ankle (induction: 6-Lead of KM6L) 11: chest (V1) 12: chest (V2) 13: chest (V3) 14: chest (V4) 15: chest (V5) 16: chest (V6) 17: abdomen 18: palm * No unit as this is a user-entered item. Fixed to 2 for SakuraOne/KardiaMobile. For KardiaMobile 6L, 2 for 1 lead and 3 for 6 leads. 11–18 are for HCG-8010T only |
| 0x1082 | Symptom | 0x0000 | None | Inclusive OR of all items corresponding to selected symptoms is saved to Value. The input text for symptoms is saved in DATA_TEXT if 'others' is selected. Bitwise inclusive OR of the following symptoms. 0x01: none 0x02: others 0x04: chest pain 0x08: dizziness 0x10: fatigue 0x20: nausea 0x40: palpitation 0x80: shortness of breath * If the value is 0, it will be considered as 'no input'. * No unit as this is a user-entered item. |
| 0x1083 | Immediately preceding intensity of exercise | 0x0000 | None | 0 : not entered 1 : high 2 : medium 3 : low 4 : resting * No unit as this is a user-entered item. |
| 0x1084 | Note (for ECG recording) | 0x0000 | None | Note by user when ECG waveform is recorded. * No unit as this is a user-entered item. |
| 0x1085 | ECG waveform graph flipped upside down | 0x0000 | None | 0 : no flipping 1 : flip upside down * No flipping other than 1 The value that is set in the details of waveform graph. The default value is 0. |
| 0x1087 | ECG waveform graph with filter settings | 0x0000 | None | 0 : no digital filtering process (Original) 1 : reserve 2 : with digital filtering process (Enhanced) The value that is set in the details of waveform graph. For SakuraOne/KardiaMobile, the default value is the one set during measurement on a filter-type device. |
HEM-7530T_J3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x1002 | Average heart rate | 0xF0A0 | bpm | |
| 0x1012 | File type | 0x0000 | None | 0 : no digital filtering process (Original) 1 : reserve 2 : digital filtering process (Enhanced) 3: for development validation Device setting information on measurement for SakuraOne/HCG-8010T. |
| 0x1013 | Analysis result type and Analysis result | 0x0000 | None | The type number of analysis result is saved on Value while the text of analysis result is saved on DATA_TEXT. 200/201/207/214/218/296/297/298/299 for SakuraOne/KardiaMobile/KardiaMobile 6L. For details, please refer to the 'ECG Analysis Result Type Details' sheet. |
| 0x1014 | ECG monitoring data classification of UUID | 0x0000 | None | UUID (string) generated by AliveCorSDK for every monitoring data |
| 0x1015 | Kardia AI lib version | 0x0000 | None | Library version for ECG analysis of AliveCorSDK (obtained from ACKEcgEvaluation) |
| 0x1016 | ECG sampling rate | 0xF0E0 | Hz | Frequency of sampling when recording ECG waveform Fixed to 125 for HCG-8010T |
| 0x1017 | Commercial power frequency | 0xF0E0 | Hz | 50 : 50Hz 60 : 60Hz Device setting information during recording |
| 0x1018 | Setting for number of leads | 0x0000 | None | 1 : single lead 6 : six leads Fixed to 1 for SakuraOne/KardiaMobile/HCG-8010T |
| 0x1019 | Maximum recording time | 0xF050 | Second | 30/60/120/180/240/300 Device setting information at the time of recording (maximum number of seconds for recording) Fixed to 30 for HCG-8010T |
| 0x101A | Actual recording time | 0xF050 | Second | Number of seconds for actual recording of ECG waveform For HCG-8010T, measurement time is calculated from the number of data entries |
| 0x101B | File path of waveform data (originalPath) | 0x0000 | None | File path for waveform data (atc file) in the filter type of Original. Shared by iOS and Android, the file path is set directly under the directory for saving atc files. |
| 0x101C | File path of waveform data (enhancedPath) | 0x0000 | None | File path for waveform data (atc file) in the filter type of Enhanced. Shared by iOS and Android, the file path is set directly under the directory for saving atc files. |
| 0x101D | ECG recording device type | 0x0000 | None | Device type of ECG recording 0 : Not set 1 : SakuraOne 2 : KardiaMobile 3 : KardiaMobile 6L 4: HCG-8010T |
| 0x102F | Average heart rate (for displaying) | 0xF0A0 | bpm | No limitation for SakuraOne/KardiaMobile. * If the value is outside the supported range: -2147483648 (INT_MIN) |
| 0x1081 | Measurement part | 0x0000 | None | 0 : Not entered (unknown) 1 : chest 2 : wrist 3 : wrist and ankle (induction: 6-Lead of KM6L) 11: chest (V1) 12: chest (V2) 13: chest (V3) 14: chest (V4) 15: chest (V5) 16: chest (V6) 17: abdomen 18: palm * No unit as this is a user-entered item. Fixed to 2 for SakuraOne/KardiaMobile. For KardiaMobile 6L, 2 for 1 lead and 3 for 6 leads. 11–18 are for HCG-8010T only |
| 0x1082 | Symptom | 0x0000 | None | Inclusive OR of all items corresponding to selected symptoms is saved to Value. The input text for symptoms is saved in DATA_TEXT if 'others' is selected. Bitwise inclusive OR of the following symptoms. 0x01: none 0x02: others 0x04: chest pain 0x08: dizziness 0x10: fatigue 0x20: nausea 0x40: palpitation 0x80: shortness of breath * If the value is 0, it will be considered as 'no input'. * No unit as this is a user-entered item. |
| 0x1083 | Immediately preceding intensity of exercise | 0x0000 | None | 0 : not entered 1 : high 2 : medium 3 : low 4 : resting * No unit as this is a user-entered item. |
| 0x1084 | Note (for ECG recording) | 0x0000 | None | Note by user when ECG waveform is recorded. * No unit as this is a user-entered item. |
| 0x1085 | ECG waveform graph flipped upside down | 0x0000 | None | 0 : no flipping 1 : flip upside down * No flipping other than 1 The value that is set in the details of waveform graph. The default value is 0. |
| 0x1087 | ECG waveform graph with filter settings | 0x0000 | None | 0 : no digital filtering process (Original) 1 : reserve 2 : with digital filtering process (Enhanced) The value that is set in the details of waveform graph. For SakuraOne/KardiaMobile, the default value is the one set during measurement on a filter-type device. |
HEM-7530T_JT3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x1002 | Average heart rate | 0xF0A0 | bpm | |
| 0x1012 | File type | 0x0000 | None | 0 : no digital filtering process (Original) 1 : reserve 2 : digital filtering process (Enhanced) 3: for development validation Device setting information on measurement for SakuraOne/HCG-8010T. |
| 0x1013 | Analysis result type and Analysis result | 0x0000 | None | The type number of analysis result is saved on Value while the text of analysis result is saved on DATA_TEXT. 200/201/207/214/218/296/297/298/299 for SakuraOne/KardiaMobile/KardiaMobile 6L. For details, please refer to the 'ECG Analysis Result Type Details' sheet. |
| 0x1014 | ECG monitoring data classification of UUID | 0x0000 | None | UUID (string) generated by AliveCorSDK for every monitoring data |
| 0x1015 | Kardia AI lib version | 0x0000 | None | Library version for ECG analysis of AliveCorSDK (obtained from ACKEcgEvaluation) |
| 0x1016 | ECG sampling rate | 0xF0E0 | Hz | Frequency of sampling when recording ECG waveform Fixed to 125 for HCG-8010T |
| 0x1017 | Commercial power frequency | 0xF0E0 | Hz | 50 : 50Hz 60 : 60Hz Device setting information during recording |
| 0x1018 | Setting for number of leads | 0x0000 | None | 1 : single lead 6 : six leads Fixed to 1 for SakuraOne/KardiaMobile/HCG-8010T |
| 0x1019 | Maximum recording time | 0xF050 | Second | 30/60/120/180/240/300 Device setting information at the time of recording (maximum number of seconds for recording) Fixed to 30 for HCG-8010T |
| 0x101A | Actual recording time | 0xF050 | Second | Number of seconds for actual recording of ECG waveform For HCG-8010T, measurement time is calculated from the number of data entries |
| 0x101B | File path of waveform data (originalPath) | 0x0000 | None | File path for waveform data (atc file) in the filter type of Original. Shared by iOS and Android, the file path is set directly under the directory for saving atc files. |
| 0x101C | File path of waveform data (enhancedPath) | 0x0000 | None | File path for waveform data (atc file) in the filter type of Enhanced. Shared by iOS and Android, the file path is set directly under the directory for saving atc files. |
| 0x101D | ECG recording device type | 0x0000 | None | Device type of ECG recording 0 : Not set 1 : SakuraOne 2 : KardiaMobile 3 : KardiaMobile 6L 4: HCG-8010T |
| 0x102F | Average heart rate (for displaying) | 0xF0A0 | bpm | No limitation for SakuraOne/KardiaMobile. * If the value is outside the supported range: -2147483648 (INT_MIN) |
| 0x1081 | Measurement part | 0x0000 | None | 0 : Not entered (unknown) 1 : chest 2 : wrist 3 : wrist and ankle (induction: 6-Lead of KM6L) 11: chest (V1) 12: chest (V2) 13: chest (V3) 14: chest (V4) 15: chest (V5) 16: chest (V6) 17: abdomen 18: palm * No unit as this is a user-entered item. Fixed to 2 for SakuraOne/KardiaMobile. For KardiaMobile 6L, 2 for 1 lead and 3 for 6 leads. 11–18 are for HCG-8010T only |
| 0x1082 | Symptom | 0x0000 | None | Inclusive OR of all items corresponding to selected symptoms is saved to Value. The input text for symptoms is saved in DATA_TEXT if 'others' is selected. Bitwise inclusive OR of the following symptoms. 0x01: none 0x02: others 0x04: chest pain 0x08: dizziness 0x10: fatigue 0x20: nausea 0x40: palpitation 0x80: shortness of breath * If the value is 0, it will be considered as 'no input'. * No unit as this is a user-entered item. |
| 0x1083 | Immediately preceding intensity of exercise | 0x0000 | None | 0 : not entered 1 : high 2 : medium 3 : low 4 : resting * No unit as this is a user-entered item. |
| 0x1084 | Note (for ECG recording) | 0x0000 | None | Note by user when ECG waveform is recorded. * No unit as this is a user-entered item. |
| 0x1085 | ECG waveform graph flipped upside down | 0x0000 | None | 0 : no flipping 1 : flip upside down * No flipping other than 1 The value that is set in the details of waveform graph. The default value is 0. |
| 0x1087 | ECG waveform graph with filter settings | 0x0000 | None | 0 : no digital filtering process (Original) 1 : reserve 2 : with digital filtering process (Enhanced) The value that is set in the details of waveform graph. For SakuraOne/KardiaMobile, the default value is the one set during measurement on a filter-type device. |
HEM-7600T
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
HEM-7600T-E
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
HEM-7600T-SH3BK
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
HEM-7600T2-JF
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
HEM-7600T_W
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
HEM-7600T_W-SH3W
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
HEM-7600T_W-Z
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
HEM-8630T-SH
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0004 | Posture (Result of Heart Zone Indicator) | 0x0000 | None | 0 : Heart Zone Indicator OFF 1 : Heart Zone OK 2 : Heart Zone NG (too high) 3 : Heart Zone NG (too low) -2147483648(INT_MIN): Invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-8732K-SH
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-8732T-SH
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
HEM-9601T-J3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0004 | Posture (Result of Heart Zone Indicator) | 0x0000 | None | 0: Heart Zone Indicator OFF 1: Heart Zone OK 2: Heart Zone NG (too high) 3: Heart Zone NG (too low) -2147483648(INT_MIN): Invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0026 | Measurement mode | 0x0000 | None | refer to 'Measurement Mode' |
HEM-9601T2-BR3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0004 | Posture (Result of Heart Zone Indicator) | 0x0000 | None | 0: Heart Zone Indicator OFF 1: Heart Zone OK 2: Heart Zone NG (too high) 3: Heart Zone NG (too low) -2147483648(INT_MIN): Invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0026 | Measurement mode | 0x0000 | None | refer to 'Measurement Mode' |
HEM-9601T_E3
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0004 | Posture (Result of Heart Zone Indicator) | 0x0000 | None | 0: Heart Zone Indicator OFF 1: Heart Zone OK 2: Heart Zone NG (too high) 3: Heart Zone NG (too low) -2147483648(INT_MIN): Invalid |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0026 | Measurement mode | 0x0000 | None | refer to 'Measurement Mode' |
HEM-9700T
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x0001 | Systolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0002 | Diastolic blood pressure value | 0x5010 | mmHg | -2147483648 (INT_MIN): invalid |
| 0x0003 | Pulse rate | 0xF0A0 | bpm | -2147483648 (INT_MIN): invalid |
| 0x0005 | Room temperature | 0x3000 | ℃ | |
| 0x0006 | Irregular pulse rate detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0007 | Body movement detection flag | 0x0000 | None | 0: not detected 1: detected -2147483648(INT_MIN): Invalid |
| 0x0008 | Cuff Wrap Guide check | 0x0000 | None | 0: loosely wrapped 1: securely wrapped |
| 0x000A | Consecutive measurement count | 0x0000 | None | 0: Single measurement 1: 1st TruRead measurement 2: 2nd TruRead measurement 3: 3rd TruRead measurement Fixed to 0 if TruRead function is disabled. |
| 0x0026 | Measurement mode | 0x0000 | None | refer to 'Measurement Mode' |
HCG-8010T1-J
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x1000 | Maximum heart rate | 0xF0A0 | bpm | |
| 0x1001 | Minimum heart rate | 0xF0A0 | bpm | |
| 0x1002 | Average heart rate | 0xF0A0 | bpm | |
| 0x1003 | Average value of instant heart rate | 0xF0A0 | bpm | |
| 0x1010 | Number of 125 Hz ECG waveform data | 0x0000 | None | |
| 0x1011 | 125 Hz ECG waveform data | 0x0000 | None | +1023 ~ -1024 [*9.757uV] |
| 0x1012 | File type | 0x0000 | None | 0 : no digital filtering process (Original) 1 : reserve 2 : digital filtering process (Enhanced) 3: for development validation Device setting information on measurement for SakuraOne/HCG-8010T. |
| 0x1013 | Analysis result type and Analysis result | 0x0000 | None | "The type number of analysis result is saved on Value while the text of analysis result is saved on DATA_TEXT. 200/201/207/214/218/296/297/298/299 for SakuraOne/KardiaMobile/KardiaMobile 6L. For details, please refer to the 'ECG Analysis Result Type Details' sheet. |
| 0x1014 | ECG monitoring data classification of UUID | 0x0000 | None | UUID (string) generated by AliveCorSDK for every monitoring data |
| 0x1016 | ECG sampling rate | 0xF0E0 | Hz | Frequency of sampling when recording ECG waveform Fixed to 125 for HCG-8010T |
| 0x1018 | Setting for number of leads | 0x0000 | None | 1 : single lead 6 : six leads Fixed to 1 for SakuraOne/KardiaMobile/HCG-8010T |
| 0x1019 | Maximum recording time | 0xF050 | Second | 30/60/120/180/240/300 Device setting information at the time of recording (maximum number of seconds for recording) Fixed to 30 for HCG-8010T |
| 0x101A | Actual recording time | 0xF050 | Second | Number of seconds for actual recording of ECG waveform For HCG-8010T, measurement time is calculated from the number of data entries |
| 0x101D | ECG recording device type | 0x0000 | None | Device type of ECG recording 0 : Not set 1 : SakuraOne 2 : KardiaMobile 3 : KardiaMobile 6L 4: HCG-8010T |
| 0x101F | Arrhythmia event information (for storage) | 0x0000 | None | |
| 0x1020 | Arrhythmia event information flag (Pause) | 0x0000 | None | 0:false, 1:true |
| 0x1021 | Arrhythmia event information flag (V-RUN) | 0x0000 | None | 0:false, 1:true |
| 0x1022 | Arrhythmia event information flag (Couplet) | 0x0000 | None | 0:false, 1:true |
| 0x1023 | Arrhythmia event information flag (VPC) | 0x0000 | None | 0:false, 1:true |
| 0x1024 | Arrhythmia event information flag (S-RUN) | 0x0000 | None | 0:false, 1:true |
| 0x1025 | Arrhythmia event information flag (Tachy) | 0x0000 | None | 0:false, 1:true |
| 0x1026 | Arrhythmia event information flag (Brady) | 0x0000 | None | 0:false, 1:true |
| 0x1027 | Arrhythmia event information flag (SVPC) | 0x0000 | None | 0:false, 1:true |
| 0x1028 | Arrhythmia event information flag (ST-T abnormality) | 0x0000 | None | 0:false, 1:true |
| 0x1029 | Arrhythmia event information flag (irregular RR interval) | 0x0000 | None | 0:false, 1:true |
| 0x102B | Arrhythmia event information flag (NOISE) | 0x0000 | None | 0:false, 1:true |
| 0x102C | Arrhythmia event information flag (others) | 0x0000 | None | 0:false, 1:true |
| 0x102D | Arrhythmia event information flag (no average waveform) | 0x0000 | None | 0:false, 1:true |
| 0x102E | Arrhythmia event information flag (no RR fluctuation rate) | 0x0000 | None | 0:false, 1:true |
| 0x102F | Average heart rate (for displaying) | 0xF0A0 | bpm | No limitation for SakuraOne/KardiaMobile. * If the value is outside the supported range: -2147483648 (INT_MIN) |
| 0x1030 | Analysis result flag | 0x0000 | None | HCG-8010T only Refer to the sheet 'Supplement: HCG-8010T Analysis Result Display Specification' |
| 0x1031 | Data format version | 0x0000 | None | Identification information to check the data storage format *Unsigned 4-byte |
| 0x1032 | Event analysis module version | 0x0000 | None | Number to check the version of event analysis module *Unsigned short 2-byte |
| 0x1033 | Time post-correction flag | 0x0000 | None | Flag used for time post-correction function. 0: Time not set 1: Time set 2: Time corrected 3: Not subject to time correction |
| 0x1034 | Contact determination information | 0x0000 | None | Flag indicating the result of contact determination. *Set the result of the following bit operations. b0: Contact OK b1: Weak contact of finger electrode b2: Unstable contact of finger electrode b3: Weak contact of chest electrode b4: Unstable contact of chest electrode b5: Reserved |
| 0x1081 | Measurement part | 0x0000 | None | 0 : Not entered (unknown) 1 : chest 2 : wrist 3 : wrist and ankle (induction: 6-Lead of KM6L) 11: chest (V1) 12: chest (V2) 13: chest (V3) 14: chest (V4) 15: chest (V5) 16: chest (V6) 17: abdomen 18: palm * No unit as this is a user-entered item. Fixed to 2 for SakuraOne/KardiaMobile. For KardiaMobile 6L, 2 for 1 lead and 3 for 6 leads. 11–18 are for HCG-8010T only |
| 0x1084 | Note (for ECG recording) | 0x0000 | None | Note by user when ECG waveform is recorded. * No unit as this is a user-entered item. |
| 0x1086 | ECG waveform graph with sensitivity settings | 0x0000 | None | 1 : 5mm/mV 2 : 0mm/mV (default) 3 : 20mm/mV * 0 : not defined (process as 10mm/mV) |
| 0x1088 | Measurement status HCG-8010T only | 0x0000 | None | Selected measurement status is saved in Value, and input symptom text is saved in DATA_TEXT when 'others' is selected. 0x0000: not selected 0x0002: others 0x0004: upon waking 0x0008: after eating/drinking 0x0010: at rest 0x0020: after using toilet 0x0040: after smoking 0x0080: after bathing 0x0100: after exercise 0x0200: after drinking alcohol 0x0400: after taking medication * No unit as it is a user input item * '0x0001: none' is treated as missing number |
| 0x1089 | Measurement posture HCG-8010T only | 0x0000 | None | Selected measurement posture is saved in Value, and input symptom text is saved in DATA_TEXT when 'others' is selected. 0x0000: not selected 0x0002: others 0x0004: sitting posture 0x0008: standing posture 0x0010: supine posture * No unit as it is a user input item * '0x0001: none' is treated as missing number |
| 0x108A | Subjective symptoms For HCG-8010T | 0x0000 | None | Logical OR of selected symptom items is saved in Value, and input symptom text is saved in DATA_TEXT when 'others' is selected. Logical OR of the following symptom bits: 0x0001: none 0x0002: others 0x0004: chest pain 0x0008: dizziness 0x0010: fatigue 0x0020: nausea 0x0040: palpitation 0x0080: shortness of breath 0x0100: chest pressure 0x0200: pricking chest pain 0x0400: difficulty breathing 0x0800: lightheadedness 0x1000: fainting 0x2000: skipped heartbeat * If the value is 0, it is treated as not entered * No unit as it is a user input item * Separate symptom input indicator IDs for SakuraOne/KardiaMobile6L and HCG-8010T (may be unified in future; define item for 0x1082 as well) |
HCG-8011B-J
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x1002 | Average heart rate | 0xF0A0 | bpm | |
| 0x1012 | File type | 0x0000 | None | 0 : no digital filtering process (Original) 1 : reserve 2 : digital filtering process (Enhanced) 3: for development validation Device setting information on measurement for SakuraOne/HCG-8010T. |
| 0x1013 | Analysis result type and Analysis result | 0x0000 | None | "The type number of analysis result is saved on Value while the text of analysis result is saved on DATA_TEXT. 200/201/207/214/218/296/297/298/299 for SakuraOne/KardiaMobile/KardiaMobile 6L. For details, please refer to the 'ECG Analysis Result Type Details' sheet. |
| 0x1014 | ECG monitoring data classification of UUID | 0x0000 | None | UUID (string) generated by AliveCorSDK for every monitoring data |
| 0x1015 | Kardia AI lib version | 0x0000 | None | Library version for ECG analysis of AliveCorSDK (obtained from ACKEcgEvaluation) |
| 0x1016 | ECG sampling rate | 0xF0E0 | Hz | Frequency of sampling when recording ECG waveform Fixed to 125 for HCG-8010T |
| 0x1017 | Commercial power frequency | 0xF0E0 | Hz | 50 : 50Hz 60 : 60Hz Device setting information during recording |
| 0x1018 | Setting for number of leads | 0x0000 | None | 1 : single lead 6 : six leads Fixed to 1 for SakuraOne/KardiaMobile/HCG-8010T |
| 0x1019 | Maximum recording time | 0xF050 | Second | 30/60/120/180/240/300 Device setting information at the time of recording (maximum number of seconds for recording) Fixed to 30 for HCG-8010T |
| 0x101A | Actual recording time | 0xF050 | Second | Number of seconds for actual recording of ECG waveform For HCG-8010T, measurement time is calculated from the number of data entries |
| 0x101B | File path of waveform data (originalPath) | 0x0000 | None | File path for waveform data (atc file) in the filter type of Original. Shared by iOS and Android, the file path is set directly under the directory for saving atc files. |
| 0x101C | File path of waveform data (enhancedPath) | 0x0000 | None | File path for waveform data (atc file) in the filter type of Enhanced. Shared by iOS and Android, the file path is set directly under the directory for saving atc files. |
| 0x101D | ECG recording device type | 0x0000 | None | Device type of ECG recording 0 : Not set 1 : SakuraOne 2 : KardiaMobile 3 : KardiaMobile 6L 4: HCG-8010T |
| 0x102F | Average heart rate (for displaying) | 0xF0A0 | bpm | No limitation for SakuraOne/KardiaMobile. * If the value is outside the supported range: -2147483648 (INT_MIN) |
| 0x1081 | Measurement part | 0x0000 | None | 0 : Not entered (unknown) 1 : chest 2 : wrist 3 : wrist and ankle (induction: 6-Lead of KM6L) 11: chest (V1) 12: chest (V2) 13: chest (V3) 14: chest (V4) 15: chest (V5) 16: chest (V6) 17: abdomen 18: palm * No unit as this is a user-entered item. Fixed to 2 for SakuraOne/KardiaMobile. For KardiaMobile 6L, 2 for 1 lead and 3 for 6 leads. 11–18 are for HCG-8010T only |
| 0x1082 | Symptom | 0x0000 | None | Inclusive OR of all items corresponding to selected symptoms is saved to Value. The input text for symptoms is saved in DATA_TEXT if 'others' is selected. Bitwise inclusive OR of the following symptoms. 0x01: none 0x02: others 0x04: chest pain 0x08: dizziness 0x10: fatigue 0x20: nausea 0x40: palpitation 0x80: shortness of breath * If the value is 0, it will be considered as 'no input'. * No unit as this is a user-entered item. |
| 0x1083 | Immediately preceding intensity of exercise | 0x0000 | None | 0 : not entered 1 : high 2 : medium 3 : low 4 : resting * No unit as this is a user-entered item. |
| 0x1084 | Note (for ECG recording) | 0x0000 | None | Note by user when ECG waveform is recorded. * No unit as this is a user-entered item. |
| 0x1085 | ECG waveform graph flipped upside down | 0x0000 | None | 0 : no flipping 1 : flip upside down * No flipping other than 1 The value that is set in the details of waveform graph. The default value is 0. |
| 0x1087 | ECG waveform graph with filter settings | 0x0000 | None | 0 : no digital filtering process (Original) 1 : reserve 2 : with digital filtering process (Enhanced) The value that is set in the details of waveform graph. For SakuraOne/KardiaMobile, the default value is the one set during measurement on a filter-type device. |
HCG-8060T
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x1002 | Average heart rate | 0xF0A0 | bpm | |
| 0x1012 | File type | 0x0000 | None | 0 : no digital filtering process (Original) 1 : reserve 2 : digital filtering process (Enhanced) 3: for development validation Device setting information on measurement for SakuraOne/HCG-8010T. |
| 0x1013 | Analysis result type and Analysis result | 0x0000 | None | "The type number of analysis result is saved on Value while the text of analysis result is saved on DATA_TEXT. 200/201/207/214/218/296/297/298/299 for SakuraOne/KardiaMobile/KardiaMobile 6L. For details, please refer to the 'ECG Analysis Result Type Details' sheet. |
| 0x1014 | ECG monitoring data classification of UUID | 0x0000 | None | UUID (string) generated by AliveCorSDK for every monitoring data |
| 0x1015 | Kardia AI lib version | 0x0000 | None | Library version for ECG analysis of AliveCorSDK (obtained from ACKEcgEvaluation) |
| 0x1016 | ECG sampling rate | 0xF0E0 | Hz | Frequency of sampling when recording ECG waveform Fixed to 125 for HCG-8010T |
| 0x1017 | Commercial power frequency | 0xF0E0 | Hz | 50 : 50Hz 60 : 60Hz Device setting information during recording |
| 0x1018 | Setting for number of leads | 0x0000 | None | 1 : single lead 6 : six leads Fixed to 1 for SakuraOne/KardiaMobile/HCG-8010T |
| 0x1019 | Maximum recording time | 0xF050 | Second | 30/60/120/180/240/300 Device setting information at the time of recording (maximum number of seconds for recording) Fixed to 30 for HCG-8010T |
| 0x101A | Actual recording time | 0xF050 | Second | Number of seconds for actual recording of ECG waveform For HCG-8010T, measurement time is calculated from the number of data entries |
| 0x101B | File path of waveform data (originalPath) | 0x0000 | None | File path for waveform data (atc file) in the filter type of Original. Shared by iOS and Android, the file path is set directly under the directory for saving atc files. |
| 0x101C | File path of waveform data (enhancedPath) | 0x0000 | None | File path for waveform data (atc file) in the filter type of Enhanced. Shared by iOS and Android, the file path is set directly under the directory for saving atc files. |
| 0x101D | ECG recording device type | 0x0000 | None | Device type of ECG recording 0 : Not set 1 : SakuraOne 2 : KardiaMobile 3 : KardiaMobile 6L 4: HCG-8010T |
| 0x102F | Average heart rate (for displaying) | 0xF0A0 | bpm | No limitation for SakuraOne/KardiaMobile. * If the value is outside the supported range: -2147483648 (INT_MIN) |
| 0x1081 | Measurement part | 0x0000 | None | 0 : Not entered (unknown) 1 : chest 2 : wrist 3 : wrist and ankle (induction: 6-Lead of KM6L) 11: chest (V1) 12: chest (V2) 13: chest (V3) 14: chest (V4) 15: chest (V5) 16: chest (V6) 17: abdomen 18: palm * No unit as this is a user-entered item. Fixed to 2 for SakuraOne/KardiaMobile. For KardiaMobile 6L, 2 for 1 lead and 3 for 6 leads. 11–18 are for HCG-8010T only |
| 0x1082 | Symptom | 0x0000 | None | Inclusive OR of all items corresponding to selected symptoms is saved to Value. The input text for symptoms is saved in DATA_TEXT if 'others' is selected. Bitwise inclusive OR of the following symptoms. 0x01: none 0x02: others 0x04: chest pain 0x08: dizziness 0x10: fatigue 0x20: nausea 0x40: palpitation 0x80: shortness of breath * If the value is 0, it will be considered as 'no input'. * No unit as this is a user-entered item. |
| 0x1083 | Immediately preceding intensity of exercise | 0x0000 | None | 0 : not entered 1 : high 2 : medium 3 : low 4 : resting * No unit as this is a user-entered item. |
| 0x1084 | Note (for ECG recording) | 0x0000 | None | Note by user when ECG waveform is recorded. * No unit as this is a user-entered item. |
| 0x1085 | ECG waveform graph flipped upside down | 0x0000 | None | 0 : no flipping 1 : flip upside down * No flipping other than 1 The value that is set in the details of waveform graph. The default value is 0. |
| 0x1087 | ECG waveform graph with filter settings | 0x0000 | None | 0 : no digital filtering process (Original) 1 : reserve 2 : with digital filtering process (Enhanced) The value that is set in the details of waveform graph. For SakuraOne/KardiaMobile, the default value is the one set during measurement on a filter-type device. |
HCG-8060T_GL
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x1002 | Average heart rate | 0xF0A0 | bpm | |
| 0x1012 | File type | 0x0000 | None | 0 : no digital filtering process (Original) 1 : reserve 2 : digital filtering process (Enhanced) 3: for development validation Device setting information on measurement for SakuraOne/HCG-8010T. |
| 0x1013 | Analysis result type and Analysis result | 0x0000 | None | "The type number of analysis result is saved on Value while the text of analysis result is saved on DATA_TEXT. 200/201/207/214/218/296/297/298/299 for SakuraOne/KardiaMobile/KardiaMobile 6L. For details, please refer to the 'ECG Analysis Result Type Details' sheet. |
| 0x1014 | ECG monitoring data classification of UUID | 0x0000 | None | UUID (string) generated by AliveCorSDK for every monitoring data |
| 0x1015 | Kardia AI lib version | 0x0000 | None | Library version for ECG analysis of AliveCorSDK (obtained from ACKEcgEvaluation) |
| 0x1016 | ECG sampling rate | 0xF0E0 | Hz | Frequency of sampling when recording ECG waveform Fixed to 125 for HCG-8010T |
| 0x1017 | Commercial power frequency | 0xF0E0 | Hz | 50 : 50Hz 60 : 60Hz Device setting information during recording |
| 0x1018 | Setting for number of leads | 0x0000 | None | 1 : single lead 6 : six leads Fixed to 1 for SakuraOne/KardiaMobile/HCG-8010T |
| 0x1019 | Maximum recording time | 0xF050 | Second | 30/60/120/180/240/300 Device setting information at the time of recording (maximum number of seconds for recording) Fixed to 30 for HCG-8010T |
| 0x101A | Actual recording time | 0xF050 | Second | Number of seconds for actual recording of ECG waveform For HCG-8010T, measurement time is calculated from the number of data entries |
| 0x101B | File path of waveform data (originalPath) | 0x0000 | None | File path for waveform data (atc file) in the filter type of Original. Shared by iOS and Android, the file path is set directly under the directory for saving atc files. |
| 0x101C | File path of waveform data (enhancedPath) | 0x0000 | None | File path for waveform data (atc file) in the filter type of Enhanced. Shared by iOS and Android, the file path is set directly under the directory for saving atc files. |
| 0x101D | ECG recording device type | 0x0000 | None | Device type of ECG recording 0 : Not set 1 : SakuraOne 2 : KardiaMobile 3 : KardiaMobile 6L 4: HCG-8010T |
| 0x102F | Average heart rate (for displaying) | 0xF0A0 | bpm | No limitation for SakuraOne/KardiaMobile. * If the value is outside the supported range: -2147483648 (INT_MIN) |
| 0x1081 | Measurement part | 0x0000 | None | 0 : Not entered (unknown) 1 : chest 2 : wrist 3 : wrist and ankle (induction: 6-Lead of KM6L) 11: chest (V1) 12: chest (V2) 13: chest (V3) 14: chest (V4) 15: chest (V5) 16: chest (V6) 17: abdomen 18: palm * No unit as this is a user-entered item. Fixed to 2 for SakuraOne/KardiaMobile. For KardiaMobile 6L, 2 for 1 lead and 3 for 6 leads. 11–18 are for HCG-8010T only |
| 0x1082 | Symptom | 0x0000 | None | Inclusive OR of all items corresponding to selected symptoms is saved to Value. The input text for symptoms is saved in DATA_TEXT if 'others' is selected. Bitwise inclusive OR of the following symptoms. 0x01: none 0x02: others 0x04: chest pain 0x08: dizziness 0x10: fatigue 0x20: nausea 0x40: palpitation 0x80: shortness of breath * If the value is 0, it will be considered as 'no input'. * No unit as this is a user-entered item. |
| 0x1083 | Immediately preceding intensity of exercise | 0x0000 | None | 0 : not entered 1 : high 2 : medium 3 : low 4 : resting * No unit as this is a user-entered item. |
| 0x1084 | Note (for ECG recording) | 0x0000 | None | Note by user when ECG waveform is recorded. * No unit as this is a user-entered item. |
| 0x1085 | ECG waveform graph flipped upside down | 0x0000 | None | 0 : no flipping 1 : flip upside down * No flipping other than 1 The value that is set in the details of waveform graph. The default value is 0. |
| 0x1087 | ECG waveform graph with filter settings | 0x0000 | None | 0 : no digital filtering process (Original) 1 : reserve 2 : with digital filtering process (Enhanced) The value that is set in the details of waveform graph. For SakuraOne/KardiaMobile, the default value is the one set during measurement on a filter-type device. |
KM-6L
| ID Hex | Name | Unit ID Hex | Unit Name | Description |
|---|---|---|---|---|
| 0x1002 | Average heart rate | 0xF0A0 | bpm | |
| 0x1012 | File type | 0x0000 | None | 0 : no digital filtering process (Original) 1 : reserve 2 : digital filtering process (Enhanced) 3: for development validation Device setting information on measurement for SakuraOne/HCG-8010T. |
| 0x1013 | Analysis result type and Analysis result | 0x0000 | None | "The type number of analysis result is saved on Value while the text of analysis result is saved on DATA_TEXT. 200/201/207/214/218/296/297/298/299 for SakuraOne/KardiaMobile/KardiaMobile 6L. For details, please refer to the 'ECG Analysis Result Type Details' sheet. |
| 0x1014 | ECG monitoring data classification of UUID | 0x0000 | None | UUID (string) generated by AliveCorSDK for every monitoring data |
| 0x1015 | Kardia AI lib version | 0x0000 | None | Library version for ECG analysis of AliveCorSDK (obtained from ACKEcgEvaluation) |
| 0x1016 | ECG sampling rate | 0xF0E0 | Hz | Frequency of sampling when recording ECG waveform Fixed to 125 for HCG-8010T |
| 0x1017 | Commercial power frequency | 0xF0E0 | Hz | 50 : 50Hz 60 : 60Hz Device setting information during recording |
| 0x1018 | Setting for number of leads | 0x0000 | None | 1 : single lead 6 : six leads Fixed to 1 for SakuraOne/KardiaMobile/HCG-8010T |
| 0x1019 | Maximum recording time | 0xF050 | Second | 30/60/120/180/240/300 Device setting information at the time of recording (maximum number of seconds for recording) Fixed to 30 for HCG-8010T |
| 0x101A | Actual recording time | 0xF050 | Second | Number of seconds for actual recording of ECG waveform For HCG-8010T, measurement time is calculated from the number of data entries |
| 0x101B | File path of waveform data (originalPath) | 0x0000 | None | File path for waveform data (atc file) in the filter type of Original. Shared by iOS and Android, the file path is set directly under the directory for saving atc files. |
| 0x101C | File path of waveform data (enhancedPath) | 0x0000 | None | File path for waveform data (atc file) in the filter type of Enhanced. Shared by iOS and Android, the file path is set directly under the directory for saving atc files. |
| 0x101D | ECG recording device type | 0x0000 | None | Device type of ECG recording 0 : Not set 1 : SakuraOne 2 : KardiaMobile 3 : KardiaMobile 6L 4: HCG-8010T |
| 0x102F | Average heart rate (for displaying) | 0xF0A0 | bpm | No limitation for SakuraOne/KardiaMobile. * If the value is outside the supported range: -2147483648 (INT_MIN) |
| 0x1081 | Measurement part | 0x0000 | None | 0 : Not entered (unknown) 1 : chest 2 : wrist 3 : wrist and ankle (induction: 6-Lead of KM6L) 11: chest (V1) 12: chest (V2) 13: chest (V3) 14: chest (V4) 15: chest (V5) 16: chest (V6) 17: abdomen 18: palm * No unit as this is a user-entered item. Fixed to 2 for SakuraOne/KardiaMobile. For KardiaMobile 6L, 2 for 1 lead and 3 for 6 leads. 11–18 are for HCG-8010T only |
| 0x1082 | Symptom | 0x0000 | None | Inclusive OR of all items corresponding to selected symptoms is saved to Value. The input text for symptoms is saved in DATA_TEXT if 'others' is selected. Bitwise inclusive OR of the following symptoms. 0x01: none 0x02: others 0x04: chest pain 0x08: dizziness 0x10: fatigue 0x20: nausea 0x40: palpitation 0x80: shortness of breath * If the value is 0, it will be considered as 'no input'. * No unit as this is a user-entered item. |
| 0x1083 | Immediately preceding intensity of exercise | 0x0000 | None | 0 : not entered 1 : high 2 : medium 3 : low 4 : resting * No unit as this is a user-entered item. |
| 0x1084 | Note (for ECG recording) | 0x0000 | None | Note by user when ECG waveform is recorded. * No unit as this is a user-entered item. |
| 0x1085 | ECG waveform graph flipped upside down | 0x0000 | None | 0 : no flipping 1 : flip upside down * No flipping other than 1 The value that is set in the details of waveform graph. The default value is 0. |
| 0x1087 | ECG waveform graph with filter settings | 0x0000 | None | 0 : no digital filtering process (Original) 1 : reserve 2 : with digital filtering process (Enhanced) The value that is set in the details of waveform graph. For SakuraOne/KardiaMobile, the default value is the one set during measurement on a filter-type device. |